Frequency Analysis |
A/D converter |
24 bits of precision. |
Automatic Integrator Correction |
Allows precise measurement of low frequency vibrations down to 0.2 Hz. |
Averaging modes |
Normal, exponential, peak hold, order tracking, negative averaging, synchronous time |
Cursors Spectrum |
Single, Harmonic, Moving Harmonic, Sideband, and Time/Frequency for waveform. |
Dynamic range |
Converter has 120 dB dynamic range. |
Frequency range |
DC to 10 Hz minimum, DC to 80 kHz maximum |
Frequency Units |
Hz, CPM, Orders. |
Full-scale range |
Accelerometer input: 0-20V, Volts input -20V + 20V |
Noise floor |
Typically less than 20 μV for a 400-line spectrum at 1000 Hz maximum frequency. |
Number of averages |
5,000 in Route mode, 10,000 in Job mode. |
Resolution |
100, 200, 400, 800, 1600, 3200, 6400, or 12800 lines of resolution. True Zoom provides effective resolution of up to 300,000 lines. |
Response |
Flat to DC for non-integrated and DC-coupled signals; optional AC coupling -3 dB at 1 Hz |
Scaling |
Linear or Log, both X and Y |
Windows |
Hanning or uniform |
Data Storage Capacity |
Internal memory |
1 GB |
External memory |
SD (Secure Digital). Virtually unlimited memory with off the shelf SDHC SD cards up to 32 GB. |
Data Analysis Speed |
400 line, 1000 HZ spectrum |
67% overlap 6 avg/sec |
1600 line, 1000 Hz spectrum |
67% overlap 3 avg/sec |