R3961T
- S-Parameter Test Set
Frequency Range
300 kHz to 3.6 GHz
Test Port Impedance: 50 ohm
Directivity:
- ≥ 25 dB, 300 kHz to 5 MHz
- ≥ 35 dB, 5 MHz to 1.3 GHz
- ≥ 30 dB, 1.3 GHz to 3.6 GHz
Input Port Return Loss
- ≥ 17 dB, 300 kHz to 1.3 GHz
- ≥ 12 dB, 1.3 GHz to 3 GHz
- ≥ 10 dB, 3 GHz to 3.6 GHz
Test Port Return Loss
- ≥ 20 dB, 300 kHz to 1.3 GHz
- ≥ 16 dB, 1.3 GHz to 3 GHz
- ≥ 14 dB, 3 GHz to 3.6 GHz
ERROR CORRECTION FUNCTIONS
- Normalize
- 1-port Calibration
- 2-port Calibration
- Data Averaging
- Auto Offset Compensation
INSTRUMENT STATE FUNCTIONS
- Save: Instrument settings are stored in an internal register by pressing the Save
Register key. Measurement data, calibration data, and instrument settings are saved
on built-in floppy disk by pressing the Store key
- Recall
- Limit Line Function
PROGRAMMING FUNCTIONS
- BASIC Controller Function
- Built-in Arithmetic Functions
- Floppy Disk Drive Functions
CONNECTION WITH EXTERNAL EQUIPMENT
- Copy: Graphs displayed on the CRT and data lists can be output to a GPIB
compatible plotter or printer without requiring an external controller
- Video Plotter Output Signals: Separate signal ( DIN 8-pin ),
Composite signal
( BNC )
- GPIB Data Output and Remote Control: Based on IEEE 488
- S-parameter Test Set Control
- Parallel I/O Output: TTL level, 8-Bit output ( 2 ports ), 4-Bit
input / output
( 2 ports )
- RS-232-C: Serial output based on RS-232-C standard
- External Trigger: BNC female connector, TTL level, LOW enable
- External Reference
Frequency Input: Frequency 1, 2, 5, 10 MHz.
Connector: BNC female connector
|
|
300kHz-3.6GHz Network Analyzer with S-Parameter Test
Set
Characteristics
MEASURING FUNCTIONS
Display Channels: Two channels
Display Parameters: A/R, B/R, A/B. Includes conversion of
impedance, admittance, S-parameters. Characteristics impedance ( Z0)
input is also possible.
Format: Orthogonal Display: Log / linear amplitude, phase,
group delay, real number and imaginary number parts of complex parameters. |Z|, R, X
(for impedance conversion measurement). |Y|, G, B (for admittance conversion
measurement). Phase extension display function.
Smith Chart: Marker readout is log/linear amplitude, phase, real number part +
imaginary number part, R+jX, G+jB.
Polar Coordinate Display: Marker readout is log/linear amplitude, phase, real
number part + imaginary number part.
SIGNAL SOURCE CHARACTERISTICS
- Measuring Frequency: Range: 300 kHz to 3.6 GHz. Resolution: 1 Hz
- Output Level: Range: +20 dBm to -5 dBm. Resolution: 0.01 dB
- Output: Single. Connector: N type (Female), 50 ohm
SWEEP FUNCTIONS
Sweep Parameters: Frequency, level
Maximum Sweep Range
Frequency: 300 kHz to 3.6 GHz
Signal Level: -5 dBm to +20 dBm
Setting Range: Start / stop or center / span
Sweep Type: Can be selected from linear or logarithmic
frequency, sweep, partial and given frequency sweep, level sweep, and CW (single
frequency)
Sweep Time: 0.5 ms / point
Measuring Points: 3, 6, 11, 21, 51, 101, 201, 301, 601, 1201
points. However, the maximum displayed number of points is 601
Sweep Trigger: Continuous, hold, single, sweep, or set by
either line trigger or external trigger
Sweep Mode:
Dual Sweep: Two channels are swept in the same frequency range
Alternate Sweep: Two channels can be swept by different sweep types in different
frequency ranges
INPUT CHARACTERISTICS
- Input Terminals: 3-CH ( R, A, B )
- Dynamic Range: 100 dB ( CH A, CH B ), 30 dB ( CH R )
- Noise Level: -90 dBm ( 1 kHz bandwidth on CH A, B ). -100 dBm ( 10 Hz bandwidth
on CH A, B )
- Minimum Input to R Channel: -30 dBm ( minimum required level to ensure correct
operation )
- Maximum Input Level: 0 dBm
- Input Breakdown Level: +20 dBm
- Input Impedance: 50 ohm
- Return Loss: 20 dB
- Connector: N-type (Female) 50 ohm
- Input Crosstalk: -100 dB (300 kHz to 1 GHz). -90 dB ( 1 GHz to 3.6 GHz )
- Resolution Bandwidth: 1 kHz to 10 Hz (Variable in 1-3-10 sequence)
- Amplitude Characteristics: Measuring Range: 0 ± 100 dB (amplitude ratio).
Amplitude Resolution: 0.001 dB
- Phase Measurement - Measuring Range: ±180° (Can be displayed in ±180°
or more by display extension function)
Phase Resolution: 0.01°
Frequency Characteristics: ±5°
GROUP DELAY TIME CHARACTERISTICS
- Linear / Logarithmic Frequency Sweeps and Sweeps with Arbitrary Frequency
- Measurement Range: 1 ps to 250 s
- Group Delay Time Resolution: 1 ps
MARKER FUNCTIONS
- Marker Display: Marker readout can be converted to display values corresponding
to each measurement format
- Multi-marker: Ten independent markers can be set for each channel
- Delta Marker
- Fixed Marker
- Correction Marker
- Marker Couple
- Analysis in Arbitrary Specified Section
- Marker Search
- Marker Tracking
- Target Search
- Marker: MKR -> reference value, MKR -> Start, MKR -> Stop, MKR ->
Center, MKR -> / Span, MKR -> Center scale
- Auto Zoom
|