| Input |
120 volts (1 Amp) AC, 50/60 Hz.,Single phase, through 7 ft. - 1/2 ft., three conductor detachable line cord, terminated in three prong grounding type plug |
| Output |
0 - 3 kV AC, Continuously adjustable. Essentially sinusoidal. Frequency same as input. Maximum current 25 mA (adjustable from 0.5 to 25 mA). No transients exceeding 120% of peak value.Peak value between 1.3 and 1.5 times RMS value |
| Sensitivity |
Factory preset at 1240 volts with a 1% 120 kohm load resistor. May be reset if necessary. When properly adjusted, response time is 0.5 second or less |
| Metering |
Analog, 0 - 3 kV with 50 volt eading resolution, connected directly across the output terminations. Voltmeter input impedance exceeds 14.7 megohms |
| Accuracy |
2%of full scale, with 2% tracking |
| Failure Detection System |
Audible and visual indication of unacceptable performance and auxiliary relay contacts (SPDT) which transfer when failure is detected. Indicators remain active and conspicuous until reset manually |
| High Voltage & Return/Continuity Terminations |
High voltage three prong isolated grounding type receptacle. Alsoequipped with detachable 5 ft.(grey) high voltage and (black) return/continuity test leads |