| Specifications: |
| INPUT |
115
VAC (± 15%), 47-63 Hz, Single Phase 230 VAC (± 15%), 47-63 Hz,
Single Phase User Selectable |
| FUSE |
115/230
VAC-5 Amp |
OUTPUT RATING |
5KV AC @ 20 mA
6KV DC @ 6 mA |
OUTPUT ADJUSTMENT |
0-5KV AC (6KV DC), 10 volt/steps
0-20.00 mA AC & 6 mA DC, 0.01 mA/steps |
HIGH TRIP RANGE |
0.05-20.00 mA AC & 0.02-6.00 mA
DC
Accuracy ±(2% of setting + 0.02 mA) |
LOW TRIP RANGE |
0.00-19.99 mA AC & 0.00-5.99 mA
DC
Accuracy ±(2% of setting + 0.02mA) |
| CONTINUITY
TRIP |
1.1
Ohm maximum at .1 Amp |
| FAILURE
DETECTOR |
Audible
& Visu& displayed on LCD) meter holds breakdown
voltage and leakage current reading after failure. |
VOLTAGE DISPLAY |
3 Digits, 6.00 KV Full Scale, LCD
Display
Accuracy: Readings ± (2% of reading + 10 volts) Settings ± (2% of
reading + 5 volts) |
| CURRENT
DISPLAY |
4
Digits, 20.00 mA Full Scale, LCD Display Accuracy: Readings ± (2%
of reading + 0.02 mA) |
| DC
OUTPUT RIPPLE |
≤5%
(6 KV DC, 6 mA) |
| AC
OUTPUT WAVEFORM |
Sine
wave, Distortion² 1% |
| AC
OUTPUT FREQUENCY |
50
or 60 Hz, User Selectable |
| OUTPUT
REGULATION |
1%
of Setting + 5 Volts |
| DWELL
TIMER |
0.5-999.9
seconds in 0.1 second increments or continuous, accuracy ± (0.1% of
reading + 0.05 seconds) |
| RAMP
TIMER |
0.5-99.9
seconds in 0.1 second increments accuracy ± (0.1% of reading + 0.05
seconds) |
| REMOTE
(PLC) |
Inputs-Test,
Reset, Memory Select Outputs-Pass, Fail, Remote Alarm, Test in
Process |
| MEMORY |
Allows
storage of up to 10 different test programs |
| SECURITY |
User
selectable password to avoid unauthorized access to test setup
program |
| LINE
CORD |
Detachable
7 ft. (2.13 m) power cable terminated in a three prong grounding
plug |
| TERMINATIONS |
5 ft. (1.52 m) high voltage and
return leads with clip and a standard U.S. style (NEMA 5-15) remote
receptacle box for testing items terminated with a line cord.
International plugs also available. |
| MECHANICAL |
Bench or rack mount (2U height) with
tilt up front feet
Dimensions: (w x h x d) 17 x 3.5 x 16.5 (432 x 88 x 419mm) Weight 25
lbs. (11.34 kg) net |
| ENVIRONMENTAL |
Operating
Temperature 32° -113° F (0° - 45°C) Relative Humidity - 0 to 95% |
| CALIBRATION |
Traceable to National Institute of
Standards and Technology (NIST). Calibration controlled by software.
Adjustments are made through front panel keypad in a restricted
access calibration mode. Calibration information stored in
non-volatile memory. |