Options
(please consult your sales representative on availability)
- -01B IEEE-488 Bus Interface
- -03A Second Input Channel
|
|
Boonton 4200 Power Meter, RF Microwattmeter Features:
- Frequency Range, 100 kHz to 100 GHz (sensor dependent)
- Power Range 0.1 nW (-70 dBm) to 5 W (+37 dBm) sensor dependent
- Dual Sensor Simultaneous Power Measurements (optional)
- 20 Measurements per Second
- Stored Calibration Data for up to 8 Sensors
SPECIFICATIONS
- Frequency Range: 100 kHz to 100 GHz, sensor dependent. Refer to the Power
Meter Sensor section
- Power Range: (Display calibrated in mW, WW, nW, dBm,
and dB relative to selected reference); -70 dBm to +37 dBm (0.1 nW to 5W), sensor
dependent
Power Reference
- Source: Internal 50 MHz oscillator with Type N female connector on front panel
- Power Output: 1.00 mW
- Calibration Factors: 3.0 dB to -3.0 dB range in 0.01 dB steps
- Ranging: Autoranging plus hold on range
- Setting Time: Diode and millimeter wave sensors, typically 0.2 to 0.5s except
1.35 to 2.5s on the lowest range. Thermocouple sensors, typically 0.5 to 6s
Recorder Output
- Watt Mode: 10 volts full-scale, proportional to indicated power over each range
- dB Mode: 8 volts equivalent to 0 dBm. Sensitivity of 1 volt per 10 dB
change over the entire dB range of sensor
- Display: 4 digit LED, 3½ digit display of power, 4 digit display of dB with 0.01
dB resolution. Auxiliary analog display, uncalibrated, proportional to recorder
output
- Limits: Entered through front panel in dB only, operable in both dB and power
modes
- Annunciators: LEDs indicate mW, µW, nW, dBm or relative dB (dBr); use of channel
1 (CH1), channel 2 (CH2, option -03) and channel 3 (CH3 = CH1 - CH2 in dB); out of dB
limits; and condition of GPIB activity (LSN, ATN, REM, TLK, option -01B)
Power Sensors
Model
(Impedance)
(RF Connector) |
Frequency Range |
Dynamic Range (dBm) |
Overload Rating
(Watts)
(dBm) |
| DIODE SENSORS |
51011 (4B)
50 ohm
N(M) |
100 kHz to 12.4 GHz |
-60 to +10 |
300 mW
+25 dBm |
51012 (4C)
75 ohm
N(M) |
100 kHz to 1 GHz |
-60 to +10 |
300 mW
+25 dBm |
51012-S/4 (4C)
75 ohm
N(M) |
100 kHz to 2 GHz |
-60 to +10 |
300 mW
+25 dBm |
51013 (4E)
50 ohm
N(M) |
100 kHz to 18 GHz |
-60 to +10 |
300 mW
+25 dBm |
51075
50 ohm
N(M) |
100 kHz to 18 GHz |
-60 to +10 |
300 mW
+25 dBm |
51051 (4G)
50 ohm
APC 3.5 (M) |
1 MHz to 26.5 GHz |
-60 to +10 |
300 mW
+25 dBm |
51015 (5E)
50 ohm
N(M) |
100 kHz to 18 GHz |
-50 to +20 |
2 W AVG.
+33 dBm |
51033 (6E)
50 ohm
N(M) |
100 kHz to 18 GHz |
-40 to +30 |
2 W AVG.
+33 dBm |
51071
50 ohm
K(M) |
10 MHz to 26.5 GHz |
-60 to +10 |
300 mW
+25 dBm |
51072
50 ohm
K(M) |
30 MHz to 40 GHz |
-60 to +10 |
300 mW
+25 dBm |
| THERMOCOUPLE
SENSORS |
51100 (9E)
50 ohm
N(M) |
10 MHz to 18 GHz |
-20 to +20 |
300 mW
+25 dBm |
51101
50 ohm
N(M) |
100 kHz to 4.2 GHz |
-20 to +20 |
300 mW
+25 dBm |
51102
50 ohm
K(M) |
50 MHz to 26.5 GHz |
-20 to +20 |
300 mW
+25 dBm |
|