| Ranges |
| L |
100 pH resolution on lowest range to 100 kH full scale on highest
range. |
| R, X, and Z |
1µ ohm resolution on lowest range to 100 Mohm full scale on
highest range. |
| C |
0.001 pF resolution on lowest range to 10 F full scale on highest
range |
| D |
resolution to 0.000001 = 1 ppm |
| Q |
to 1,000,000 counts maximum reading |
| Speed |
|
DIRECT |
SORT and
PASS/FAIL |
| FAST |
~4 measurements/ second |
~9 measurements/ second |
| MEDIUM |
~2 measurements/ second |
~2 measurements/ second |
| SLOW |
~5 seconds/ measurement |
~5 seconds/ measurement |
| Test Signals |
| Frequency |
20 Hz to 150 kHz in 3023 discrete steps |
| Voltage Level |
5 mV to 1500 mV in 1 mV steps |
| Current Level |
0.1 mA to 100 mA in 0.1 mA steps |
|
|
Video Bridge Impedance Test Instrument
- Over 3000 frequencies from 20 Hz to 150 kHz
- CRT display
- Fully automatic setup mode
- Variable test voltage and current levels
- 13 bin sorting
- Remote control via optional IEEE-488 or standard RS-232
- Tape drive stores setups
- Input Protection - Protection Circuitry absorbs up to 1 joule
from a charged component.
- External Bias - +50 VDC maximum limit 100 mA
- Integration Time - Selectable from 2 ms to 600 ms
- Setting Time - Selectable from 2 ms to 1500 ms in 1 ms
steps.
- Number of Averages - Selectable from 1 to 20
- Display - Up to 6 digits selected automatically for
measurement results, dependent on measurement speed and frequency.
- Test Connection - Four terminal with passive guard.
Parameters Measured and Displayed
- Cs - Series Capacitance
- Cp - Parallel Resistance
- Ls - Series Inductance
- Lp - Parallel Inductance
- Rs - Series Resistance
- Rp - Parallel Resistance
- D - Dissipation Factor
- Q - Quality Factor
- Gs - Series Conductance
- Gp - Parallel Conductance
- Xs - Series Reactance
- Xp - Parallel Reactance
- Bs - Series Susceptance
- Bp - Parallel Susceptance
- Y - Admittance
- Z - Impedance
|