| Fisher F-130A Bulk Current Injector Probe
The Fischer Custom Communications F-130A is a broad band
high efficiency injection probe ideally suited for susceptibility testing
in accord with GM 9112P and SAE -J15447. The useable frequency range is 1
MHz to 700 MHz with an insertion loss of 6 dB or less from 4 MHz to 450
MHz and a CW input power rating of 50 watts for a duration of 15 minutes.
Clamp-on
Physical
Dimensions* |
Rated
Watts
CW |
Insertion
Loss |
ID
(A) |
OD
(B) |
Ht.
(C) |
< or =
6 dB |
< or =
10 dB |
< or =
15 dB |
< or =
20 dB |
| 32 |
94 |
64 |
50 |
4 MHz -
450 MHz |
1 MHz -
550 MHz |
700 KHz -
600 MHz |
500 KHz -
625 MHz |
* All dimensions are in millimeters |