Fluke 5520A
Fluke 5520A Fluke 5520A Multi-Product Calibrator
General Specifications
Warmup Time Twice the time since last warmed up, to a maximum of 30 minutes
Settling Time Less than 5 seconds for all functions and ranges except as noted
Standard Interfaces IEEE-488 (GPIB), RS-232
Temperature Performance

Operating: 0°C to 50°C
Calibration (tcal): 15°C to 35°C
Storage: -20°C to 70°C

Temperature Coefficient

Temperature coefficient for temperatures outside tcal ±5°C is 10% of the stated specification per °C for temperatures in the range of 0°C to 35°C. Above 35°C, the temperature coefficient is 20% of the stated specification per °C

Relative Humidity

Operating: <80% to 30°C, <70% to 40°C, <40% to 50°C
Storage: <95%, noncondensing
Note: After long periods of storage at high humidity, a drying out period (with the power on) of at least one week may be required

Altitude

Operating: 3,050m (10,000 ft) maximum
Non-operating: 12,200m (40,000 ft) maximum

Safety

Designed to comply with IEC 1010-1 (1992-1); ANSI/ISA-S82.01-1994; CAN/CSA-C22.2 No. 1010.1-92

Analog Low Isolation 20V
EMC Designed to comply with IEC 61326-1/1997
Line Power

Line Voltage (selectable): 100V, 120V, 220V, 240V
Line Frequency: 47 to 63 Hz
Line Voltage Variation: ±10% about line voltage setting
Note: For optimal 5520A performance at full dual outputs (e.g. 1000V, 20A), choose a line voltage setting that is ±7.5% from nominal

Power Consumption 600 VA
DC Voltage Specifications
Ranges

Absolute Uncertainty,
tcal ±5°C
±(ppm of output + μ
V)

Stability Resolution

MaximumBurdenCurrent1

90 Days 1 Year

24 Hours, ± 1° C ±(ppm of output + μV)

0 to 329.9999 mV 15 + 1 20 + 1 3 + 1 0.1 μV 50Ω
0 to 3.299999V 9 + 2 11 + 2 2 + 1.5 1 μV 10 mA
0 to 32.99999V 10 + 20 12 + 20 2 + 15 10 μV 10 mA
30 to 329.9999V 15 + 150 18 + 150 2.5 + 100 100 μV 5 mA
100 to 1000V 15 + 1500 18 + 1500 3 + 300 1000 μV 5 mA
Auxiliary Output (dual output mode only)2
0 to 329.999 mV 300 + 350 400 + 350 30 + 100 1 μV 5 mA
0.33 to 3.29999V 300 + 350 400 + 350 30 + 100 10 μV 5 mA
3.3 to 7V 300 + 350 400 + 350 30 + 100 100 μV 5 mA
TC Output and Measure in Linear 10 μV/°C and 1 mV/°C Modes
0 to 329.999 mV 40 + 3 50 + 3 5 + 2 0.1 μV 10Ω

1 Remote sensing is not provided. Output resistance is <5 mΩ for outputs ≥0.33V. The AUX output has an output resistance of <1Ω. TC output has an output impedance of 10Ω ±1Ω
2
Two channels of dc voltage output are provided. Note: Maximum voltage output is 1020V.

DC Current Specifications
Ranges Absolute Uncertainty, tcal ±5°C
±(ppm of output + μA)

Resolution

Compliance
Voltage

Maximum
Inductive Load

90 Days 1 Year
0 to 329.999 μA 120 + 0.02 150 + 0.02 1 nA 10V 400 μH
0 to 3.29999 mA 80 + 0.05 100 + 0.05 0.01 μA 10V 400 μH
0 to 32.9999 mA 80 + 0.25 100 + 0.25 0.1 μA 7V 400 μH
0 to 329.999 mA 80 + 2.5 100 + 2.5 1 μA 7V 400 μH
0 to 1.09999A 160 + 40 200 + 40 10 μA 6V 400 μH
1.1A to 2.99999A 300 + 40 380 + 40 10 μA 6V 400 μH
0 to 10.9999A 380 + 500 500 + 500 100 μA 4V 400 μH
11 to 20.5A1 800 + 750 1000 + 750 100 μA 4V 400 μH

1 At 20°C, full 20.5A output is available continuously for 20 minutes.

Please contact us for more information on the Fluke 5520A


Agilent
E4437B

Agilent E4437B
Yokogawa
WT1600B


Yokogawa WT1600B
Anritsu
MS2026A

Anritsu MS2026a
Megger
DLRO-200-115

Megger DLRO-200-115

Teseq
NSG437
Varian 956 Porta Test
Solar
7054-1

Chauvin Arnoux 6121
Chroma
63209
Chroma 63209
Privacy Notice Site Map Terms of Use
Copyright © 1996-2007  |  Advanced Test Equipment Rentals  |  All Rights Reserved