RLC Digibridge
- 0.2% accuracy for R, L, and C measurements
- 1 kHz and 120 Hz test frequencies
- Dual displays
- Built-in Kelvin test fixture
- Series and parallel measurement selection
- Wide measurement ranges
Measurement Modes: Measures R series or parallel; L and Q
series or parallel; C and D series or parallel. All measurement modes are pushbutton
selectable
Measurement Speed: Greater than three measurements per second,
unqualified
Test Frequencies: 1 kHz and 120 Hz. Pushbutton selectable
Displays: Two LED numerical displays with automatically
positioned decimal points and illumination of units. For RLC, five digits (99999)
and simultaneously for DQ, four digits (9999)
Applied Voltage: 0.3 V rms maximum
Ranges: Pushbutton selection with automatic front-panel
guidance
| Parameter |
Minimum Low Extension |
Basic Ranges |
Maximum High Extension |
| R: 120 Hz |
00.001 ohm |
2 ohm to 2 Mohm |
99.999 Mohm |
| R: 1 kHz |
00.001 ohm |
2 ohm to 2 Mohm |
9.9999 Mohm |
| L: 1 kHz |
0.0001 mH |
0.2 mH to 200 H |
999.99 H |
| L: 120 Hz |
00.001 mH |
2 mH to 2000 H |
9999.9 H |
| C: 1 kHz |
0.0001 nF |
0.2 nF to 200 µF |
999.99 µF |
| C: 120 Hz |
00.001 nF |
2 nF to 2000 µF |
99999 µF |
| D (with C) |
|
.0001 to 9.999 |
|
| Q (with L) |
|
00.01 to 999.9 |
|
|