Advanced Test Equipment Corporation Rentals - Sales - Service
Advanced Test Equipment Corporation
1-888-404-ATEC (2832)
tagline

      The Knowledge. The Equipment. The Solution


spacer

Home - Component Test - Meters - Impedance
Genrad 1658
Genrad 1658 RLC Digibridge
  • 0.1% accuracy for R, L, and C
  • Automatic limit comparison and binning
  • Two test frequencies (1 kHz and 120 Hz)
  • IEEE bus/handler option
  • External DC bias capability
  • Full-digit, dual displays
  • Built-in Kelvin test fixture

Measurement Parameters and Modes: Series or parallel R and Q, series or parallel L and Q, series or parallel C and D.  Continuous-repetitive, single, or averaged (set of 10) measurements; start button initiates single or averaged measurements.   Keyboard selection of these and all measurement conditions.

Main Displays (3 selections): Value display is LED-type numerical readout with automatically positioned decimal points and illumination of units; five digits for RLC (99999) and simultaneously four digits for DQ (9999). Limits display shows comparator bin limits and nominal values.  Bin No. display shows the bin assignment of the measured device.

Measurement Rates: Approximately 2, 3, and 7 measurements/second.  Keyboard selections are: "slow, medium, fast."

Test Frequencies: Keyboard selection between 2.  Actual frequencies: 120.00 Hz and 1020.0 Hz.

Ranges: Automatic ranging for best accuracy; autorange can be inhibited by keyboard selection.  Three basic ranges of 2 decades each, for each parameter.  Automatic extensions to min and max, as tabulated.

Parameter Minimum Low Extension Basic Ranges Maximum High Extension
R: 1 kHz 0.0001 ohm 2 ohm to 2MW 9.9999 Mohm
R: 120 Hz 0.0001 ohm 2 ohm to 2MW 99.999 Mohm
L: 1 kHz .00001 mH 0.2 mH to 200 H 999.99 H
L: 120 Hz 0.0001 mH 2 mH to 2000 H 9999.9 H
C: 1 kHz .00001 nF 0.2 nF to 200 µF 999.99 µF
C: 120 Hz 0.0001 nF 2 nF to 2000 µF 99999 µF
Q (with R)   .0001 to 9.999  
Q (with L)   00.01 to 999.9  
D (with C)   .0001 to 9.999  

Bias: Connector for external voltage source, on-off switch, and indicator for light.  Limit, 60 V (max).  External source requirements: ripple < 1 mV pk-pk, dynamic Z << 1W with currents of ± 50 mA pk (source and sink); external discharge circuit recommended.

Supplementary Display: Parameters, modes, overrange and underrange conditions, range held, bias on, and remote control.

Sorting: Limit comparator sorts vs a DQ limit and up to 8 pairs of RLC limits into 10 bins, conveniently defined by keyboard entries.  GO/NO GO is indicated, whether bin number of measured value is selected as main display.

Please contact us for more information on the Genrad 1658


Haefely
ECOMPACT4

Haefely Ecompact 4
CA Instruments
5001ix


California Instruments 5001ix
Voltech
PM6000

Voltech PM6000
IFI
T4026-40

IFI T4026-40

Teseq
NSG437
Varian 956 Porta Test
Solar
2654-1

Solar 2654-1
Terms & Conditions Privacy Notice Site Map Terms of Use
 