6½ Digit Multimeter
- 12 measurement functions
- 1000 V maximum input
- 15 ppm basic, dV accuracy (24 hour)
- 1000 readings per second direct to HP-IB
- HP-IB and RS-232 standard
- 512-reading memory
- SCPI commands standard
SPECIFICATIONS
DC VOLTAGE
- Input Characteristics
| Range |
Maximum
Reading (6½ digits) |
Resolution in
digits - 6½ |
Input
Resistance |
| 100 mV |
120.0000 |
100 nV |
10 Mohm or > 10 Gohm |
| 1V |
1.200000 |
1 µV |
10 Mohm or > 10 Gohm |
| 10 V |
12.00000 |
10 µV |
10 Mohm or > 10 Gohm |
| 100 V |
120.0000 |
100 µV |
10 Mohm |
| 1000 V |
1050.000 |
1 mV |
10 Mohm |
- Measurement Accuracy: ± (% or reading + % of range)
| Range |
1-Year
23° C ±5° C |
| 100 mV |
0.0050 + 0.0035 |
| 1 V |
0.0040 + 0.0007 |
| 10 V |
0.0035 + 0.0005 |
| 100 V |
0.0045 + 0.0006 |
| 1000 V |
0.0045 + 0.0010 |
AC VOLTAGE (true rms)
- Measurement Accuracy: ± (%of reading + % of range); 1 year, 23° C ± 5° C
| Frequency |
Ranges
100 mV |
Ranges
1, 10, 100 750 V |
| 3 to 5 Hz |
1.00 + 0.04 |
1.00 + 0.03 |
| 5 to 10 Hz |
0.35 + 0.04 |
0.35 + 0.03 |
| 10 Hz to 20 kHz |
0.05 + 0.04 |
0.06 + 0.03 |
| 20 to 50 kHz |
0.12 + 0.04 |
0.12 + 0.05 |
| 50 to 100 kHz |
0.60 + 0.08 |
0.60 + 0.08 |
| 100 to 300 kHz |
4.00 + 0.50 |
4.00 + 0.50 |
- Input Resistance: 1 MW ±2%, in parallel with 100 pf
Frequency and Period
- Range: 3 Hz to 300 kHz (333 ms to 3.33 µs)
- Resolution: 10 µHz to 1 Hz
- Other Measurement Functions: Continuity, Diode Test, Ratio dc: dc, Limit Test
- Math Functions: NULL, Min/Max/Avg, dB, dBm, Limit Test
- Memory: 512-reading internal storage
- Standard Programming Languages: SCPI, HP 3478A AND Fluke 8840A/42A
- Computer Interface: HP-IB and RS-232C standard
RESISTANCE: (2-wire ohm, 4-wire ohm)
- Input Characteristics
| Range |
Maximum reading
(6½) |
Resolution in digits
6½ |
| 100 ohm |
120.0000 |
100 µohm |
| 1 kohm |
1.200000 |
1 mohm |
| 10 kohm |
12.00000 |
10 mohm |
| 100 kohm |
120.0000 |
100 mohm |
| 1 Mohm |
1.200000 |
1ohm |
| 10 Mohm |
12.00000 |
10 ohm |
| 100 Mohm |
120.0000 |
100 ohm |
- Measurement Accuracy: ± (% of reading + % of range)
Specs are for 4-wire ohm or 2-wire ohm using Math Null
| Range |
1-Year
23° C ±5° C |
| 100 ohm |
0.010 + 0.004 |
| 1 kohm |
0.010 + 0.001 |
| 10 kohm |
0.010 + 0.001 |
| 100 kohm |
0.010 + 0.001 |
| 1 Mohm |
0.010 + 0.001 |
| 10 Mohm |
0.040 + 0.001 |
| 100 Mohm |
0.800 + 0.010 |
DC CURRENT
- Measurement Accuracy: ± (% of reading + % of range)
| Range |
1-Year
23° C ±5° C |
| 10 mA |
0.050 + 0.020 |
| 100 mA |
0.050 + 0.005 |
| 1 A |
0.100 + 0.010 |
| 3 A |
0.120 + 0.020 |
- Input Protection: Externally accessible 3A 250V fuse; internal 7A 500V fuse
AC CURRENT (true rms)
- Measurement Accuracy: ± (% of reading + % of range); 1 year, 23° C ± 5° C
| Frequency |
Ranges
1A |
Ranges
3A |
| 3 to 5 Hz |
1.00 + 0.04 |
1.10 + 0.06 |
| 5 to 10 Hz |
0.30 + 0.04 |
0.35 + 0.06 |
| 10 Hz to 5 kHz |
0.10 + 0.04 |
0.15 + 0.06 |
|