5½ Digit Multimeter
- Five Functions
- Up to 100 nanovolt resolution
- Electronic calibration
- HP-IB interface
- Front/rear terminals
Specifications
DC Voltage
- Input Characteristics
| Range |
Maximum Reading (5½
digit) |
| 30 mV |
± 30.3099 mV |
| 300 mV |
± 303.099 mV |
| 3 V |
± 3.03099 V |
| 30 V |
± 30.3099 V |
| 300 V |
± 303.099 V |
- Input Resistance:
30 mV, 300 mV, 3 V ranges: > 1010 ohm
30 V, 300 V ranges: 10 Mohm ± 1%
- Maximum Input Voltage (nondestructive): Hi to Lo: 303 Vrms or 450 V peak; Hi or
Lo to Earth Ground: 500 V peak
AC Voltage (true rms responding)
- Input Characteristics
| Range |
Maximum Reading (5½
digit) |
| 300 mV |
303.099 mV |
| 3 V |
3.03099 V |
| 30 V |
30.3099 V |
| 300 V |
303.099 V |
- Input Impedance: 1 Mohm ± 1% shunted by < 60 pF
- Frequency: 20 Hz to 300 kHz
Resistance (2-wire ohm and 4-wire ohm)
- Input Characteristics
| Range |
Maximum
Reading (5½ digit) |
| 30 ohm |
30.3099 ohm |
| 300 ohm |
303.099 ohm |
| 3 kohm |
3.03099 kohm |
| 30 kohm |
30.3099 kohm |
| 300 kohm |
303.099 kohm |
| 3 Mohm |
3.03099 Mohm |
| 30 Mohm |
30.3099 Mohm |
DC Current
- Input Characteristics
| Range |
Maximum Reading (5½
digit) |
| 300 mA |
303.099 mA |
| 3 A |
3.03099 A |
- Maximum Input (nondestructive): 3 A from < 250 V source: fuse protected
AC Current (true rms responding)
- Input Characteristics
| Range |
Maximum Reading (5½
digit) |
| 300 mA |
303.099 mA |
| 3 A |
3.03099 A |
|