Accessories Available:

HP / Agilent 16091A Coaxial Test Fixture
HP / Agilent 16092A Spring Clip Test Fixture
HP / Agilent 16093A/16093B Binding Post Test Fixture
HP / Agilent 16904A Probe Fixture
|
|
RF Impedance Analyzer
Specifications:
Parameters Measured: |Z| - q,
|Y| - q, vector reflection coefficient - q , R - X, G -B, vector reflection coefficient
(real/imaginary components), L - R . G . D . Q, C - R . G . D . Q.
Display: 4½ digit, max. display 19999 counts.
Deviation Measurement (deviation from stored reference):
- 19999 to + 19999 counts; -1999.9 to + 19999.9%.
Measuring Signal:
- Frequency Range: 1 MHz to 1000 MHz
- Frequency Step: 100 kHz, 1 to 500 MHz; 200 kHz, 500 to 1000 MHz
Signal Level: (into 50 ohms): -20 ± 3 dBm.
Frequency Control: Spot and sweep.
Measuring Mode:
- Spot Measurement: At specific frequency ( or dc bias).
- Swept Measurement: Manual or automatic sweep from start to stop frequency (
or dc bias) at step frequency ( or dc bias) rate in linear or logarithmic form.
Auto Calibration: Automatic error compensation referenced
to connected terminations (0 ohms, 50 ohms, 0 S), 51 frequencies including start and stop
frequencies.
Electrical Length Compensation: Automatic compensation
for electrical length of test fixtures (range: 0 to 99.99 cm).
Internal dc Bias:
- Voltage Range: -40 to +40 V, 10 mV step
- Bias Control: spot and swept
External dc Bias:
- Voltage Range: -40 to +40
- Max Allowable Current: 100 mA
Key Status Memory: Two sets of measuring conditions can
be stored and recalled at any time.
Ranging: Auto/Range hold.
Trigger: Internal, External, or Manual
HP-IB Data Output and Remote Control: Standard |