HP / Agilent 4194A Impedance / Gain-Phase Analyzer
The HP / Agilent 4194A Impedance / Gain-Phase Analyzer
is an integrated solution for efficient measurement and analysis or
go/no-go testing of components and circuits. Detailed impedance and
transmission characteristics including secondary parameter derivations can
be simply and quickly evaluated or tested. The HP / Agilent 4194A can
contribute to improving engineering productivity and reducing test cost.
The analyzer is flexible and has a wide measurement capabilities in both
impedance and transmission measurements, plus it is fully programmable
using Auto Sequence Programming (ASP). Desired measurements and
computations, including graphics analysis, can be programmed simply by
storing front panel keystroke operations, allowing you to customize
measurement, computation and analysis functions. The HP / Agilent 4194A
also features high accuracy and error elimination functions to ensure
reliable measurements.
Features:
- High Accuracy and Wide Range:
Impedance Measurement:
100Hz to 40MHz, 0.1mΩ to 1.6MΩ,
0.17%
10 kHz to 100 MHz, 0.1Ω to 1MΩ,
1.5% when used with the HP / Agilent 41941A/B
Gain-Phase Measurement:
10 Hz to 100MHz, -107dBm to +15dBm, 0.1dB
resolution
- Flexible Measurement, Computation and Analysis
Capabilities on a Color Graphic Display
- Fully Programmable
Specifications:
Impedance Measurements
- Measurement Parameters: |Z|, |Y|, θ, R, X, G, B,
L, C, D, Q. 20 parameter combinations are available.
- Test Frequency: 100 Hz - 40 MHz (cable length: 0m),
100Hz - 15 MHz (cable length: 1m), 1mHz resolution
- OSC Level: 10 mV - 1Vrms (≤10MHz), 10mV -
0.5Vrms (> 10MHz), 3 digit resolution
- DC Bias:0 - +40V, 10mV resolution
- Measurement terminal: 4 - terminal pair configuration
- Measurement Range and Max. Resolution:
Measurement
Parameter |
Range |
Max.
Resolution |
| |Z|,
R, X |
10mΩ
to 100MΩ |
100μΩ |
| |Y|,
G, B |
10nS
to 100S |
1nS |
| θ |
+ 180° |
0.01° |
| L |
1nH
to 100kH |
10pH |
| C |
10fF
to 0.1F |
0.1fF |
| D |
0.001
to 10 |
0.0001 |
| Q |
0.1
to 1000 |
0.1 |
- Basic Measurement Accuracy: 0.17%
Gain-Phase Measurements
- Measurement Parameters: Tch/Rch (dB, Linear Ratio),
Tch, Rch (V, dBm, dBV), θ(degree, rad),
τ
Tch = Test Channel, Rch = Reference Channel, τ = Group Delay
- Measurement Frequency: 10 Hz - 100MHz, 1mHz
resolution
- APerture Frequency Range (Group Delay Measurements):
0.5% - 100% of frequency span
- OSC Level: -65dBm - + 15dBm, 0.1dB resolution
Measurement Range:
Tch/Rch: 0 - +120dB, 0.001dB resolution
Tch/Rch: -107dBm - -5dBm (+dB Attenuator); -87dBm - +15dBm (20dB
Attenuator); 0.001dB resolution
θ: +180° (can
display phase continuously with the phase scale expansion function),
0.01° resolution
τ: 0.1ns - 1s, 0.1ns resolution
- Basic Measurement Accuracy:
Tch/Rch: 0.1dB, 0.5°
Tch/Rch: 0.35dBm
θ: 0.5°
|