Accessory:
- HP / Agilent 16117B Low-Noise Test Leads (1 m, 2 Clips)
Available Test Fixtures:
|
|
HP / Agilent 4339A High-Resistance Meter
- Wide measurement range: 1 x 103 ohm to 1.6 x 1016ohm
- Stable test fixtures: resistivity cell, component test fixture
- High-speed measurement: 10 ms
- Test sequence programming
- Resistivity calculations
- Grounded DUT measurement
Specifications
Measurement Functions
- Measurement Parameters: R (dc resistance), I (dc current), rs
(surface resistivity), rv (volume resistivity)
- Mathematical Functions: Deviation and percent deviation
- Ranging: Auto and manual
- Trigger: Internal, external, manual, and HP-IB
- Delay Time: 0 to 9999 ms in 1 ms steps
- Test Cable Lengths: 2 m maximum
- Measurement Time: Short, medium, and long
- Averaging: 1 to 256
- Test Voltage: 0 to 1000 Vdc, 100 mV steps @ 0 to 200 V, 1 V steps @ 200 to 1000 V
- Maximum Current: 10 mA @ ≤ 100 V, 5 mA @ ≤ 250V, 2 mA @ ≤500 V, 1 mA @ ≤ 1 kV
- Current Compliance Setting: 0.5 mA, 1 mA, 2 mA, 5 mA, 10 mA
Measurement Range/Accuracy
| Parameter |
Measurement Range |
Basic Accuracy |
| I |
60 fA to 100 µA |
± 0.4% |
| R (ohm) |
1 x 103 to 1.6 x 1016 |
± 0.6% |
Measurement Time:
| Mode |
Time (typical) |
| Short |
10 ms |
| Medium |
30 ms |
| Long |
1850 ms |
Correction Function: Zero OPEN
Other Functions: Test Sequence Program; Comparator Function;
Contact Check Function; Save/recall; Continuous Memory Capability; HP-IB Interface;
Handler Interface |