Features
- Automatic calibration and zeroing
- Frequency entry instead of calibration factor
- Ten pre-loaded sensor calibration factor versus frequency
tables
- Selectable resolution to 0.001 dB
- Offset entry in dB
- Duty-cycle entry for a convenient peak power representation of
the measured average power
- Ten store/recall registers
- HP-IB is standard
- Analog meter is a standard feature
|
|
Power Meter
Specifications
- Frequency Range: 100kHz to 110GHz, sensor dependent
- Power Range: -70dBm to +44dBm (100 WP to 25W), sensor dependent
- Power Sensors: Compatible with all HP 8480 series power sensors
- Dynamic Range: 50 dB in 10 dB steps
- Display Units: Absolute: watts, dBm; relative: percent, dB
- Resolution: Selectable resolution of 0.1, 0.01, and 0.001 dB in logarithmic mode; or 1% and 0.01 of full
scale in linear mode
Accuracy
- Instrumentation: ±0.02 dB or ±0.5%
- In REL Mode: ±0.02 dB or ±0.5% within measurement range; ±0.04 dB or 1% outside measurement range
- Zero Set: ±0.5% of full scale on most sensitive range
Power Reference
- Power Output: 100 mW, factory-set to ±0.7% traceable to US
National Institute of Standards and Technology
|