| Home - Oscilloscopes |
| Hitachi V1100A |
|
 |
|
100 MHz Quad Channel Oscilloscope
|
| CRT Readout Function |
| Direct Measurement through CH 1 input |
ACV RMS, ACV Relative, DCV Deflection, Frequency
Counter |
| Cursor Measurement |
Ground reference voltage, Voltage difference, Voltage difference
ratio, Time difference, Phase difference, Frequency |
| Ground reference display |
GND level for CH 1 and CH 2 displayed with waveforms |
| Setting Conditions Display |
Vertical Deflection: Scale Value for CH1 and CH2, 10:1 probe
display |
| Horizontal Deflection: Scale Value for time base A and B,
UNCAL, MAG |
| Other: A and B trigger source, X - Y, Delay Time |
| User- Defined Comment Display |
1 line up to 30 characters |
|
| Vertical Deflection |
| Sensitivity |
CH 1 and CH 2 |
5 mV/div to 5 V/div (1 mV/div , 2mV/div when using magnifier) |
| CH 3 and CH 4 |
0.1 V/div, 0.5 V/div |
| Bandwidth |
DC to 100 MHz, DC to 15 MHz when using x5 magnifier |
| Rise Time |
3.5 ns |
| Input Coupling |
CH 1 and CH 2: AC, GND, DC |
| CH 3 and CH 4: AC, DC |
| Input Impedance |
1 M ohm approximately 28 pF |
| Display Mode |
CH 1, CH 2 (normal or invert). ALT. CHOP, ADD (CH 1 and
CH 2), QUAD (ALT, CHOP) |
|
| X - Y Operation (CH
1: X-axis, CH 2: Y-axis) |
| Sensitivity |
5 mV/div to 5 V/div |
| 1 mV/div, 2 mV/div when using x5 magnifier |
| Bandwidth |
DC to 2 MHz |
|
| Horizontal Deflection |
| A Sweep Time Range |
20 ns/div to 0.5 s/div |
| 2 ns/div to 10 ns/div when using x10 amplifier |
| B Sweep Time Range |
20 ns/div to 50 ms/div |
| 2 ns/div to 10 ns/div when using x10 amplifier |
| Display Mode |
A, INTEN ( A intensified by B), ALT ( A intensified by B and B), B. |
|
|
|