Keytek EMCPro Advanced EMC Test System for Compliance Testing
Configurations for the Keytek EMC Pro:
| Model
PRO-SURGE |
| Surge
per IEC 61000-4-5. Will do part of UL1449 |
| Voltage
Waveform: |
1.2/50µs |
| Peak
Voltage: |
250V
- 6.6kV ±5%, 12Ω Mode
250V - 6.0kV ±5%, 2Ω mode |
| Peak
Current: |
125A
- 3.3kA ±10% |
| Additional
10Ω Resistor: |
Software
selectable |
| Repetition
Rate: |
Up
to 4 per minute |
| Open-circuit
Voltage: |
Front
time: 1.2µs ±30%
Duration: 50µs ±20%
Undershoot: ≤30% |
| Line
sync accuracy: |
±15%,
50-277VAC |
| Model
PRO-EFT |
| EFT
per IEC 61000-4-4 and ANSI |
| Voltage
Waveform: |
5/50ns
±30% |
| Peak
Voltage: |
250V
- 4.4kV ±5% |
| Burst
Period: |
300ms
±10% |
| Burst
Duration: |
15ms
±20% |
| Frequency: |
1-100kHz
in 0.5kHz steps, ±10% |
| DC
Blocking Capacitor: |
10nF
(internal) |
| Model
PRO-RING* |
| Surge
Ring Wave per ANSI and UL. Will do part of 61000-4-12. |
| Voltage
Waveform: |
100kHz
damped cosine |
| Peak
Voltage: |
250
- 6.6kV ±5% |
| Repetition
Rate: |
<4/minute
at 6kV, faster at lower voltages |
| Open-circuit
Voltage: |
Rise
Time: 0.5µs ±30% |
| Short-circuit
Current: |
Vp/Ip:
12Ω ±3Ω or 30Ω ±8Ω software selectable. |
| *Pro-Telecom
and Pro-Ring can not be installed together in the same unit. |
| Surge
Waveform Monitoring |
| Lines
Monitored: Monitors are automatically switched to match the
generator coupling mode. |
| Open-circuit
Voltage: |
1000:1
±10% |
| Short-circuit
CurrentAttenuation: |
200:1
±7% |
| Model
PRO-HPOWER |
| Power
Frequency Magnetic Field per 61000-4-8 |
| Field
Frequency: |
50Hz/60Hz |
| Field
Amplitude: |
0.5-4A/m,
in 0.25A steps, ±10% (with CM-HCOIL) up to 100A/m with the optional
external HPOWER-EXT |
| AC
Source: |
Internal |
| Resolution: |
0.25A
minimum |
| Coil
Factor: |
0.65
to 1.00 |
| Coil
Resistance: |
0.05Ω
maximum |
| Model
PRO-HPULSE |
| Pulse
Magnetic Field per IEC 61000-4-9 |
| Field
Pulse: |
8/20µs |
| Field
Amplitude: |
100A/m
-1000A/m, ±10% |
| Resolution: |
5A/m |
| Coil
Factor: |
0.65
to 1.00 |
| Model
PRO-PQF |
| Dips
and Interrupts per IEC 61000-4-11 |
| Dips: |
70%,
40% |
| Interrupts: |
0%
(short and open) |
| Transition
Time: |
1µs
- 5µs |
| Inrush: |
Minimum: |
250Amps
@ 100-120V |
| |
Minimum: |
500Amps
@220-240V |
| AC
Voltage: |
50-250VAC,
50-60Hz |
| AC
Current: |
16A
Max |
| PQF
Sync Output: |
5V
signal occurs at each Dip or
interrupt transition |
| Dip
Duration: |
10msec
min at 60Hz
2msec
min at 50Hz |
| PQF
Waveform Monitoring |
| Voltage
Input Connection: |
Fixed,
L1 to L2 |
| Voltage
Attenuation: |
100:1
±5% |
| Current
Input Connection: |
Fixed,
L1 |
| Peak
Current: |
Minimum
500A inrush into 1700µF |
| Current
Attenuation: |
200:1
±5% |
| Model
PRO-BASE |
| System
Voltage: |
90-240VAC,
50-60Hz |
| Integrated
EUT Mains Coupler/Decoupler |
| AC
Voltage: |
1
phase, 50-250VAC, 50/60Hz |
| AC
Current: |
16A
max |
| DC
Voltage: |
100VDC
max |
| DC
Current: |
10A
max |
| Frequency: |
50/60Hz |
| EUT
Connectors: |
Nema,
British, Schuko |
| Control
Interface: |
RS232
Fiber-optic |
| Safety
Features: |
External
Interlock for users
Interlock for CCL connector
External stop input |
Environmental Operating Conditions |
| Temperature: |
15-40°C |
| Humidity: |
10-75%,
non-condensing |
| Altitude: |
8000
ft. max. |
|