Specifications
FM Measurement
- Carrier Frequency Range: 4 to 1000 MHz
- Deviation Range: Five ranges with full-scale indications of 5, 15, 50, 150 and
500 kHz. Positive or negative deviation indication selected by a switch
AM Measurement
- Carrier Frequency Range: 4 to 350 MHz
- Mod. Depth Range: Two ranges with full-scale indications of 30% and 100% (maximum
usable reading: 95%). Peak and trough indications selected by a switch
RF Input
- Sensitivity:
Less than 20mV between 4 and 250 MHz
Less than 50mV between 250 and 500 MHz
Less than 100mV between 500 and 1000 MHz
- Maximum Input: 3V RMS (200mW)
- Input Impedance: Nominally 50 ohm
Local Oscillator
- Variable Frequency Operation: The internal oscillator covers two ranges: 5.5 to
11 MHz and 22 to 44 Mc/s, harmonics being used for other local oscillator
frequencies. Provision is made for connecting an external oscillator with an output
of approximately 20mV.
IF Output
- Frequency: 1.5 MHz
- Amplitude: Between approximately 250 and 750mV e.m.f.
LF Output
- Frequency Range: 30 Hz to 150 kHz with switchable 15 kHz low-pass filter on
f.m. 30 Hz to 50 kHz on a.m.
- De-emphasis: Switchable to 0, 50 or 75µsec