Panametrics 22DLHR Ultrasonic Thickness Meter
Specifications
- Thickness Measurement Range: 0.005-10" or 0.1-200mm. Thickness range
also depends on material and transducer type
- Thickness Resolution:
| 0.01" |
LOW |
0.1mm |
| 0.001" |
STANDARD |
0.01mm |
| 0.0001" |
HIGH |
0.001mm |
- Measurement Rate: 2 per second
- Velocity Calibration Range: 0.0300-0.551 in./uS; 0.762-13.999
mm/uS
- Differential Mode: Permits measurements relative to a preset nominal value
- Differential Mode Setpoint Range: 0-20" or 0-200mm
- Measurement Range (Differential Mode): 0.005-10" or 0.13-200mm
- Data Logger Functions: Unit will store, recall, clear and transmit via RS-232 up
to 1000 thickness readings
- Output Format: Data is transmitted in a serial ASCII string
- Baud Rate: Selectable: 150, 300, 600, 1200, 1800, 2400, 3600, 4800, 7200, or 9600
- Output Connector: Ruggedized 6 pin modular telephone type
- Power Requirements: 6VDC (supplied from internal batteries) or 100-130VAC,
50-60Hz supplied to external charger provides unit operating power for all conditions of
battery charge state including recharging
- Battery: 6V Rechargeable NiCad battery pack
Recommendations
| Material |
Transducer (1)
Type |
Transducer
Frequency |
Thickness (2)
Range |
| Polyethylene / Polypropylene |
M116 |
20 MHz |
.005 - .125"
0.125 - 3mm |
| Polystyrene / Pet / Acrylic |
M116 |
20 MHz |
.005 - .500"
0.125 - 12.5mm |
| Mild Steel |
M116 |
20 MHz |
.020 - 1.5"
0.375 - 37.5mm |
| Glass |
M116 |
20 MHz |
.020 - 1.5"
0.375 - 37.5mm |
(1) For a more complete summary of thickness gage applications and contact
transducer recommendations, contact your Sales Representative.
(2) Note: Ranges are approximate and will vary depending on material sound
velocity, surface condition, and attenuation. |