36DL Plus Ultrasonic Thickness Gage
- Standard Measurement Mode:
Time interval from a precision delay after the excitation pulse to the first echo.
- Thru-Paint Echo-to-Echo Mode:
Time interval between two successive backwall echoes for elimination of paint or coating using dual or single element transducers. Manual Mode or Automatic Mode.
- Thickness Measurement Range:
0.020-20 inches (0.5-500mm).
- Material Velocity Range:
0.0300-0.5511 inch/µS.(0.762-13.999mm/µS).
- Manual Gain Adjust:
Allows the gain to be manually increased or decreased in 1dB steps.
- Automatic Probe Recognition:
The 36DL PLUS automatically adjusts internal parameters and corrects for V-path.
- Zero Compensation Mode:
Provides zero offset compensation for temperature variations in the transducer material, and for transducer wear.
- Alarm Mode:
Programmable Hi-Low Alarm set points with audible and visual indicators.
- File Structures:
Data can be stored in 8 standard or custom application-specific file structures.
- Reports:
On-gage reporting of: Summary with statistics, Min/Max with locations, and File Comparison. On-screen comparison of current and previous readings.
- Interface Program:
16-bit or 32-bit Windows Interface Program compatible with Windows 3.x, 95 and NT.
- Datalogger and RS-232:
The 36DL PLUS identifies, stores, recalls,clears, and transmits thickness readings, waveform images, and gage setup information via the RS-232 Serial Port. Baud Rate, Word Length, Stop Bits, and Parity are adjustable from the keypad.
- Maximum # of Stored Values:
Up to 95,000 thickness readings and/or 4000 waveforms, or combination of both.
- Stored Data Documentation:
Each saved thickness reading or waveform is fully documented with measurement status flags and a setup number which identifies parameters such as velocity, transducer gain, etc.
|