Applications:
- Testing of LCR Components
- Frequency Response Characterization of Components
- Component Screening
- Material Testing
- Quality Assurance Testing
- Measuring Dielectric Constant Using Standard Test
Cell
Features:
- 7 Digit Measurement Resolution
- Auto Ranging
- Test Setup and Measurement Data Storage
- Four-Terminal Kelvin Connections
- IEEE-488, RS-232, Handler, and Parallel Printer
Interfaces
- Graphical and Tabular Display of Swept Frequency,
Voltage and Current Measurements
- Sequence Testing of up 6 Individual Tests
- Load Correction
- Built-in Automatic Calibration Routine
There are various options and available for the QuadTech
7600, including the 7000-07 Chip Component Test
Fixture. |
|
Precision LCR Meter
Specifications:
- Measured Parameters: Any two of 14 parameters
measured and displayed simultaneously, user selectable
| Parameter |
Measurement Range |
| Cs, Cp |
00000.01 fF to 9.999999 F |
| Ls, Lp |
0000.001 nH to 99.99999 H |
| D |
.0000001 to 99.99999 |
| Q |
.0000001 to 999999.9 |
| |Z|, Rs, Rp, ESR, Xs |
000.0001 mohms to 99.99999 Mohms |
| |Y|, Gp, Bp |
00000.01 µS to 9.999999 MS |
| Phase Angle |
-180.0000 to +179.9999 degrees |
- Test Frequency: 10 Hz to 2 MHz,
continuous
* Resolution: 0.1 Hz from 10 Hz to 10 kHz, 5 digits>10kHz
* Accuracy: +/-(0.25% + 0.02 Hz)
- Measurement Speed:
* Basic: 25 measurements/second
* Medium: 8 measurements/second
* Slow: 1 measurements/second
- Ranging: Automatic, Range Hold or user selectable
- Trigger: Internal (auto), External (RS-232, IEEE-488.2 or
Handler interfaces) and Manual
- AC Test Signal
Voltage:
* 20 mV to 1.0 V (open circuit) 500 kHz - 1 MHz in 5 mV steps
* 20 mV to 0.5 V (open circuit) >1 MHz in 5 mV steps Current
Current: 250 µA to 100 mA (short circuit) in 50 µA steps
|