Hand-held Configurator /
Calibrator
Intrinsically safe interface for use with
Rosemount smart transmitters.
Features
- Common interface for all Rosemount SMART FAMILY® instruments
- Digital communications via the HART protocol without
interruption of transmitter output signal
- Easy to operate
- Microprocessor-based electronics
- Large 4 line x 20 character display
Specifications
- Memory: Nonvolatile memory retains contents when unit
is powered down, so long as the battery source is supplying power.
- Power Supply: Five AA 1.5 V batteries (rechargeable
optional)
- Temperature Limits:
- Operating: 32 to 122 °F (0 to 50 °C)
- Storage Limits: -4 to 158 °F (-20 to 70 °C)
- Humidity Limits: 0 to 95% RH
- Hazardous Locations Certifications:
- CENELEC/BASEEFA Intrinsic Safety Certification
- Factory Mutual (FM) Intrinsic Safety and Non-incendive
Approval
- Canadian Standards Association (CSA) Intrinsic
Safety Approval
- Standards Association of Australia (SAA)
Intrinsic Safety Certification
- Display: 4-line liquid crystal display with
20-character line width
- Keyboard: Complete alphanumeric keyboard, six
dedicated function keys, and four software-defined keys
- Weight: 2 lbs (0.9 kg) including batteries
|