Teseq / Schaffner NSG 4070-75 Signal Generator and Compact Immunity Test System.

The NSG 4070, successor of the NSG 2070, is a multifunctional EMC immunity test system. Its large frequency range from 9kHz to 1GHz and its modular set-up using internal or external amplifiers enable a large variety of applications including tests according to IEC 61000-4-6, various BCI applications as well as signal generator and power meter for test systems as per IEC 61000-4-3, IEC 61000-4-20, IEC 61000-4-21 and many other applications.

The powerful and easy to use firmware makes the NSG 4070 independent from an external PC and control software, however it can also be remote controlled for system operation. A state-of-the-art data transfer of test and measurement data for documentation is provided by USB stick to be plugged into the front panel.

Features

  • Integrated signal generator 9kHz to 1GHz
  • 4-channel power meter 9kHz to 1GHzAn advanced combination of signal generator and compact immunity test system.
  • Integrated power amplifier module for different
  • frequency ranges up to 1GHz and up to 75W
  • nominal output power
  • Multiple EUT monitoring options
  • Internal, menu-based control software
  • 5,7“ TFT color display
  • Optional remote control software
Generator Specifications

RF:  
Frequency range 9 kHz-1 GHz
Resolution 0,1 Hz
Reference frequency 20 MHz internal, 10 MHz Reference output
RF Level:  
Level range -60 dBm to +10 dBm
Resolution 0,1 dB
Settling time < 120 µs (10kHz bandwidth)
NF-Generator:  
Frequency range 1 Hz - 50 kHz
Frequency resolution 1 Hz
Amplitude modulation:  
Modulation depth 0-100%
Modulation frequency range 1 Hz - 50 kHz
Pulse modulation:  
Rise / fall time (10% / 90%) < 1 µs
Modulation frequency range 1 Hz - 50 kHz / 50% duty cycle
External modulation:  
Delay time < 1 µs / 1800
Period min. 20 µs
Pulse width min. 10 µs

Power Amplifier Specifications

Nominal output power: 75W
Frequency range: 150 kHz - 230 MHz
Input impedance: 50 Ohm
Output impedance: 50 Ohm
Input return loss min.: 10 dB
Output return loss: nom. min. 9,5 dB (0 dB without damage )
Gain: min. 50 dB
Gain flatness: max. +/- 3 dB
Saturated output power: min. 48,75 dBm
Max. input power linear without damage: < -3 dBm max. +10 dBm
2nd harmonic distortion at nominal output power: typ. < -35 dBc
3rd harmonic distortion at nominal output power: typ. < -18 dBc

Please contact us for more information on the Teseq / Schaffner NSG 4070-75

Featured Test Equipment:

Rent an Haefely ECompact 4 Transient Immunity Test Set

Haefely ECOMPACT4
Transient Immunity Test

Voltech PM6000
Voltech PM6000
Power Analyzer



IFI T4026-40
TWT Amplifier

Varian 956 Porta Test
Teseq NSG 438
ESD Gun

Solar 2654-1
Solar 2654
Transient Generator

Forms Terms & Conditions Privacy Notice Home Page Terms of Use
Copyright © 1996-2010  |  Advanced Test Equipment Rentals  |  All Rights Reserved