Deflection Controls
Display Mode - Norm and Dc Modes - Vertical Collector
Current - Horizontal Collector Volts - Horizontal Base Volts - Step Generator Mode -
Horizontal Mode
Vertical Deflection Factor - Collector
Current: 2 nA/div to 2 A/div, 28 steps (0.2 nA/div to 0.2 A/div with x10
magnification).
Horizontal Deflection Factor:
Characteristics
Device Lead Selection - Switch provides six
different lead configurations. Three positions for Emitter Grounded measurements
provide Step Gen, Open (or Ext), and Short base terminal connections. Two positions
for Base Grounded measurements provide Step Gen and Open (or Ext) emitter terminal
connections. One position provides for Emitter Base Breakdown or leakage
measurements up to 25 Volts.
Left-Right Switch
Variable Voltage Supply - Continuously
variable bias supply from -12 V to + 12 V. |
|
Storage and Nonstorage Curve Tracer
System
- Test Two-Terminal and Three-Terminal Discrete Semiconductors
- Storage Capability (Model D1)
- Power Capability Up to 100 Watts
Collector Supply
Modes/Polarity - Norm: Ac (at line frequency);
positive or negative full-wave rectified ac. DC: positive or negative
dc.
Series Resistance - From 0.12 ohms to 8 M ohms in
14 steps. Peak Power Limit Setting: 100 W, 30 W, 9 W 2.3 W, 0.6 W, 0.15 W.
Safety Interlock
| Voltage |
| Range |
6.5 V |
25 V |
100 V |
400 V |
1600 V |
| Max Continuous Peak Current |
10 A |
2.5 A |
0.6 A |
0.15 A |
0.04 A |
Step Generator
Current Mode - Step/Offset Amplitude Range:
5 nA/step (with x0.1 Mult) to 200 mA/step.
Voltage Mode - Step/Offset Amplitude Range:
5 mV/step (with x0.1 Multi) to 2 V/step
Step Rates - Selectable at x1, x2, or x4
line frequency.
Step Family - Repetitive or single family
Number of Steps - Selectable from 1 to 10
full-amplitude steps
|