| Input Voltage |
208VAC 3 phase |
| Recommended
Min. Service Amperes |
1
phase 40 amperes
3 phase 30 amperes |
| Temperature
Range |
-20°F
to +350°F
-29°C to +177°C |
Humidity
Range
(Limited by a +40°F (+4°C) minimum
dew point and a max. dry bulb of
+185°F (+85°C)) |
20%
to 95% |
| Humidity
Water Operating at Full Capacity |
0.5
GPH
1.9 LPH |
Temperature
Pulldown From +75°F (+24°C)
-20°F (-29°C) |
50 minutes |
Temperature
Heat Up From +75°F (+24°C)
+200°F (+93°C)
+350°F (+177°C) |
18 minutes
30 minutes |
Capacity
for Holding Watts Live Load
0°F (-18°C)
-20°F (-29°C)
95% R.H.
50% R.H. |
200 watts
50 watts
200 watts
500 watts |
**CHAMBER REQUIRES EXTERNAL WATER SUPPLY FOR COOLING**
The
above specifications are based on 60 hertz operation and +27°C
(+80°F) ambient air.
For 50 hertz operation, refrigeration and air flow performance are
approximately 17% less. |