Options:
- 20mA - Low Current Version
- I-150 - Clamp-on 150A CT
- I-1000 - Clamp-on 1000A CT
|
|
AC-DC Wideband Digital
Power Analyzer
Range/Resolution Table:
| True RMS
Voltage |
True RMS Current |
| .2000A |
2.000A |
20.00A |
| 150.00V |
30.00W |
300.0W |
3000W |
| 300.0V |
60.00W |
600.0W |
6000W |
| 600.0V |
120.00W |
1200.0W |
12000W |
| |
True WATTS |
Specifications:
AC/DC Voltage
& AC/DC Current (True RMS)
- Range & Accuracy:
- DC & 40 Hz to 5 kHz, ± 0.1% of reading ± 6
digits
- 5 kHz to 15 kHz, ± 0.5% of reading ± 6 digits
- 15 kHz to 20 kHz, ± 0.75% of reading ± 6 digits
- to 50 kHz with typical 1% error per 10 kHz
- Crest Factor Response: 50:1 for minimum RMS input,
linearly decreasing to 2.5:1 for full scale RMS
- Minimum Input: 5% of range
- Maximum Input: 600V DC or RMS AC, 1500V peak
- Maximum Common Mode: 1500V peak, neutral to earth
- Peak Indicator: Illuminates at 2.5 times full scale
- Maximum Input: 35A peak, 20A DC or RMS; 100A DC or RMS
for 16ms without damage
- Overrange: 150% of full scale for DC, up to maximum
input
WATTS (True Power-EI cos )
- Range & Accuracy:
- DC & 40 Hz to 5 kHz, ± 0.25% of reading ± 6
digits
- 5 kHz to 10 kHz, ± 0.5% of reading ± 6 0.5% of
range
- 10 kHz to 20 kHz, ± 1% of reading ± 1% of range
- 20 kHz to 50 kHz, additional ± 1% per 10 kHz above
20 kHz
- Power Factor Response: Unity to zero leading or lagging
- Power Factor Resolution: (.001 pf) zero to 1.000 pf
(1000 counts)
General Specifications:
- Displays: dual 4½ digit
- Load Connection: 4 Terminal heavy duty binding posts
- Size: 9.25" L x 5" W x 8.5" H (23.5 cm x 21.6 cm
x 6.4 cm)
- Weight: 5 lbs (2.3 kg)
|