| AC/DC Voltage & AC/DC Current (True RMS) |
| Range & Accuracy |
DC & 40 Hz to 5 kHz, ± 0.1% of reading ± 6 digits |
| 5 kHz to 15 kHz, ± 0.5% of reading ± 6 digits |
| 15 kHz to 20 kHz, ± 0.75% of reading ± 6 digits to 50 kHz with typical 1% error per 10 kHz |
| Crest Factor Response |
50:1 for minimum RMS input, linearly decreasing to 2.5:1 for full scale RMS |
| Minimum Input |
5% of range |
| Maximum Input |
600V DC or RMS AC, 1500V peak |
| Maximum Common Mode |
1500V peak, neutral to earth |
| Peak Indicator |
Illuminates at 2.5 times full scale |
| Maximum Input |
35A peak, 20A DC or RMS; 100A DC or RMS for 16ms without damage |
| Overrange |
150% of full scale for DC, up to maximum input |
| WATTS (True Power-EI COS Ø) |
| Range & Accuracy |
DC & 40 Hz to 5 kHz, ± 0.25% of reading ± 6 digits |
| 5 kHz to 10 kHz, ± 0.5% of reading ± 0.5% of range |
| 10 kHz to 20 kHz, ± 1% of reading ± 1% of range (2 A range only) |
| 20 kHz to 50 kHz, additional ± 1% per 10 kHz above 20 kHz |
| Power Factor Response |
Unity to zero leading or lagging |
| Power Factor Resolution |
(.001 pf) zero to 1.000 pf (1000 counts) |
| Displays |
Dual 4½ digit |
| Load Connection |
4 Terminal heavy duty binding posts |
| Size |
9.25" L x 5" W x 8.5" H (23.5 cm x 21.6 cm x 6.4 cm) |
| Weight |
5 lbs (2.3 kg) |