Sensitive Wide Range Survey Meter
- 1 mR/hr Most Sensitive Full Scale Range
- Integrate and Rate Capability
- Energy Corrected from 6 keV to 2 MeV
Specifications
- Radiation Detected: Alpha above 3.4 meV, beta above 70 keV, gamma and X-rays
above 6 KeV
- Operating Ranges: Rate: 12 overlapping ranges: 0-1, 3, 10, 30, 100 and 300 mR/h
and R/hr; 0-10, 30, 100 and 300 µSv/h, 1 and 3 mSv/h; 0-10, 30, 100 and 300 mSv/h, 1 and
3 Sv/h. Integrate: 6 overlapping ranges: 0-1, 3, 10, 30, 100, and 300 mR; 0-10, 30,
100, and 300 µSv, 1 and 3 mSv
- Response Time: 90% of final reading within 8 sec. on 1 mR/hr range, 3 secs on 3
mR/hr range decreasing to less than 1 sec. on all other ranges
- Detector: Unsealed air ionization chamber with 1.1 mg/cm² mylar end window.
Supplied with protective cover of tissue equivalent plastic 300 mg/cm²
- Display: 7.9 cm (3 1/8") taut band meter
- Battery Complement: Two 1.5 volt "D" cells and four 22.5 volt No. 505
batteries
|