Advanced Test Equipment Corporation Rentals - Sales - Service
Advanced Test Equipment Corporation
1-888-404-ATEC (2832)
tagline

      The Knowledge. The Equipment. The Solution


spacer

Home - Signal Conditioners
Vishay 2210

Vishay 2210 Signal Conditioning Amplifier System
Vishay 2210 Signal Conditioning Amplifier System Plug-In

 

 

Constant-Voltage Excitation:
Range:
0.50 to 15.0 Vdc at 85 mA max.

Noise:
+ 100 microvolts + 0.002% p-p dc to 20 kHz.

Line Regulation:
+ 200 microvolts + 0.01 % max for line voltage change of 10% from nominal.

Load Regulation:
+ 200 microvolts + 0.01% max for load variation of 10% to 90% of full load.

Stability:
+ 0.01%/deg C or 100 microvolts/deg C, whichever is greater.

Remote Sense:
Error <0.0005%/ohms of lead resistance.

Monitoring:
Front-panel monitoring jacks.

Isolation:
Isolated from power ground and output common; floats with guard.

Constant-Current Excitation:
Range:
0.50 to 15.0 mA dc or 1.00 to 30.0 mA dc. Compliance voltage 0.50 to 16.0V.

Noise:
(1 microamps + 10 microvolts) p-p; dc to 20 kHz.

Line Regulation:
+ 1 microamps + 0.01% max for line voltage change of + 10% from nominal.

Load Regulation:
+ 1 microamps + 0.01% max for 100% load change.

Stability:
+ 0.01%/deg C or 1 microamp/deg C, whichever is greater.

Monitoring:
Front-panel monitoring jacks; 10 mV/mA.

Isolation:
Isolated from power ground and output common; floats with Guard.

Balance:
Method:
Electronically injected automatic balance.

Range:
+ 15 000 microstrain (7.5 mV/V) RTI (X2 with internal jumper).

Resolution:
0.50/microstrain RTI (X2 with internal jumper).

Balance Time:
4 seconds typical; 8 seconds max.

Accuracy:
+ 2 mV RTO; + 2 microstrain RTI.

Balance Trim:
+ 375 microstrain (188 microvolts/V) RTI.

Storage:
Digital with battery backup. Battery life 3-5 years.

Activation:
Activated by front-panel switch or by optically isolated remote switch or low TTL level.

Calibration:
Four internal shunt calibration resistors, + 0.1% tolerance:
    174.8K    1000 microstrain (0.50 mV/V)    350 ohms bridge;
    874.8K    200 microstrain (0.10 mV/V)    350 ohms bridge;
    59.94K    1000 microstrain (0.50 mV/V)    120 ohms bridge.

Activated by front-panel switch, or by optically isolated remote contact closure or low TTL level.
Internal selector switches for selection of two-point unipolar, bipolar, or two-point double shunt calibration circuits.
Calibration resistors plug into fixed terminals (no soldering).

Vishay 2210 Signal Conditioning Amplifier System
  • Isolated constant-voltage/constant-current excitation
  • Guarded input structure with + 350V common-mode capability, + 10V and tape outputs
  • Automatic wide-range bridge balance
  • Four-pole Bessel low-pass filter.
  • Constant-voltage or constant-current excitation, calibration configuration, and other optional operating modes
Specifications
All references to microstrain assume a gage factor of 2.00.
All specifications nominal or typical at +23 deg C unless noted.

Input:
Input Impedance:
dc-coupled: 22 megohms shunted by 250 pF.
ac-coupled: 1.1 microfarad in series with 20 kilohms; low frequency cutoff (3 dB) 8 Hz nom.

Source Current:
+ 10 nA typical; + 20 nA maximum.

Configuration:
2- to 10-wire plus guard shield accepts quarter-, half-, or full-bridge strain gage or transducer inputs. Internal half-bridge, dummy 350 ohm and dummy 120 ohm completion gages, remote sense and four-wire calibration capability provided. 1000 ohms completion capability also provided. Accepts inputs from ground-referenced or isolated devices.

Differential Input:
Maximum differential input voltage of + 50 Vdc or peak ac.

Common-Mode Input:
Maximum common-mode input voltage of + 350 Vdc or peak ac.

Guard Impedance:
Greater than 250 kilohms to output common; greater than 1000 megohms to power and rack ground.

Amplifier:
Gain:
1 to 3300; continuously variable; direct reading. Gain steps X1, X10, X100, X300; with 10-turn counting knob, X1 to X11. Accuracy + 0.5%.

Linearity:
+ 0.01% of full scale at dc.

Frequency Response:
dc to 100 kHz: 3 + 0.2 dB at all gain settings and full output. dc to 50 kHz: 0.5 dB max at all gain settings and full output.

Gain Step vs Frequency Response (3 dB):
X300 100 kHz
X100 120 kHz
X10 135 kHz
X1 240 kHz

Slew Rate:
6.3 V/microsec min at all gain settings.
Noise:
(350 ohm source impedance, dc-coupled)
Referred-to-input (RTI):
1 microvolt 0.1 Hz to 10 Hz p-p; 2 microvolts 0.1 Hz to 100 Hz p-p
3 microvolts 0.1 Hz to 100 kHz rms
Referred-to-Output (RTO):
Output related noise is a function of the setting of the gain multiplier potentiometer. Refer to the graph below for noise referred-to-output.


Zero Stability:
+ 2 microvolts RTI, + 200 microvolts RTO at constant temp.

Temperature Coefficient of Zero:
+ l microvolt/deg C RTI, + 100 microvolt/deg C RTO; -10 deg to 60 deg C.

Common-Mode Rejection:
Gain CMR (dB)
X1 82
X10 102
X100 122
X300 135

Common-Mode Voltage:
+ 350 Vdc or peak ac, max operating.

Standard Output:
+ 10V at 10 mA max;

Tape Output:
1.0 Vrms at 10 mA max; or

Output ac-coupled:
+ 10V at 10 mA max (7 Hz, 3 dB).

Output Monitor:
+ 10V standard monitored via front-panel jacks.

Output Isolation:
>1000 megohms from power and rack ground.

Output Protection:
Protected against continuous short.

Capacitive Loading:
Up to 0.15 microfarad.

Low-Pass Filter:
Four-pole Bessel low-pass filter with selectable 3 dB bandwidths of 1 Hz, 10 Hz, 100 Hz, 1 kHz and 10 kHz.

Please contact us for more information on the Vishay 2210


Haefely
ECOMPACT4

Haefely Ecompact 4
CA Instruments
5001ix


California Instruments 5001ix
Voltech
PM6000

Voltech PM6000
IFI
T4026-40

IFI T4026-40

Teseq
NSG437
Varian 956 Porta Test
Solar
2654-1

Solar 2654-1
Terms & Conditions Privacy Notice Site Map Terms of Use
 