Microwave Radio Test Set
POWER METER
- Absolute Measurement Accuracy
Within ±0.5dB,when using the 75ohms power head over the range 10 to
300MHz or the 50ohms power head over the range 100MHz to 11.7GHz,
including the effects of ancillary equipment .
- Accuracy of Power Meter Alone
Within ±0.2dB
- RF band (5052)with low pass filter :
500 to 700MHz
700 to 1000MHz
2.11 to 2 .18GHz
3.65 to 4 .25GHz
5 .8 to 6 .5GHz
10.6 to 11 .7GHz
- IF band (75ohms) with low pass filter :
70MHz and 74 .1MHz
- Range
+10 to -30dBm without external pads or amplifiers .
- Resolution
0.010
DIGITAL VOLTMETER
- Ranges
±0.1 to ±100VDC and ±100 to ±999VDC
- Accuracy
±0.3V on 100V range and ±3V on 999V range .
SPECTRUM ANALYZER
- General Requirement
Provides a visual indication of the spectra of the total IF
bands of fo ±32MHz with a dynamic range of over 70dB .
- Scan Width Selection IF Bands
(fo =70MHz &74 .1 MI-17. fo is centered on display)
±1MHz
±5MHz
±15MHz
±32MHz
- Dynamic Range
70dB (±1MHz)
60dB (all other scan widths)
10dB sensitivity/cm of deflection
RF SWEEP GENERATOR
- CW Mode
2GHz band :2 .11 to 2 .18GHz (option 003)
4GHz band :3 .65 to 4 .25GHz
6GHz band :5 .8 to 6 .5GHz
11GHz band :10.6 to 11 .7GHz (option 001)
- Frequency Setting Accuracy
Within ±1 x 10-5
IF SWEEP GENERATOR
- CW Mode
70MHz and 74 .1MHz
- Frequency Setting Accuracy
Within ±2 x 10-5
FREQUENCY COUNTER
- Frequency Range
14 to 300MHz directly and without prescaling.
- Input Network
At least 1Mohms shunted by less than 20 pF (14 to 135MHz) and 50ohms (14
to 300MHz)switchable .
- Input Level
15mV to SVrms without need for level adjustment.
- Accuracy
±1 count ±Time Base oscillator stability.
|