Advanced Test Equipment Rentals spacer
Advanced Test Equipment Rentals
Request a Rental QuoteRequest a Sales QuoteRequest a Service QuoteContact Us
iso9001

spacer
  Home line Equipment line Manufacturers line Request line Contact Us line About Us line Events line Forms line Employment  
Home - Recorders
Astro-Med Dash 16U

Astro-Med Dash 16U Astro-Med 16 Channel Signal Conditioner/Recorder

The Dash 16u features 16 channels of universal signal conditioning, real-time data display, real-time chart recording and built-in data acquisition - all in one compact, rugged case.

High Frequency Response

Each channel in the Dash 16u has a separate A/D converter sampling at 200 kHz. Imagine – you’ll never have to worry about missing important data again!

Real-time Monitor

The Dash 16u features a 10.4-inch color display for realtime and post capture viewing of waveform data. The display can be run in place or in parallel with the chart, saving paper during setup and recording.

High Speed Memory

The Dash 16u has 16 Megabytes of memory for capturing important data. Sample rates are adjustable from 0.2 to 200,000 samples per second, ideal for transient capture as well as long-term trending.

ZIP Drive

With the integral 100 Megabyte ZIP drive you can easily upgrade software, transfer data for archive, and save complete setups.

Isolated, Universal Inputs

The Dash 16u inputs are truly universal, letting you connect almost any signal – voltage to temperature to pressure and strain – without external conditioning. Each channel is electrically isolated, eliminating crosstalk.

Specifications
Color Display

  • Type 10.4”, Active matrix color LCD (TFT)
  • Function Control menus, real-time waveforms, waveform
    review
Chart Recorder
  • Paper Direct Thermal, Z-fold pack (sheet size is 8.5” x 5.5”), top of form mark on back
  • Resolution 12dpm (300 dpi) amplitude and time axes
  • Chart Speed 1 mm/min to 50 mm/sec.
  • Dual Speed Automatic change of speed with trigger
  • Amplitude Grids 16 grids up to 200 mm wide, grid placement can
    be automatic or user determined
  • Time Marking Tri-State (x1, x10, x100) mark or either chart edge;
    Grid time lines can be synchronized to time mark;
    Selectable time mark reference (0.02 to 1 sec.)
  • Time Marking Tri-State (x1, x10, x100) mark or either chart edge;
    Grid time lines can be synchronized to time mark;
    Selectable time mark reference (0.02 to 1 sec.)
  • Channel ID Each channel labeled with channel number; Top and bottom grid values can be annotated
Data Capture and Review
  • Sample Rates 0.2 to 200,000 samples/second/channel
  • Amplitude Resolution 14 bits
  • Total Capture Memory 16 Megabytes (512 ksamp./channel)
  • Time Stamp Time and Date automatically saved with data
  • Header Information on units, range, sample rates, etc.
    saved with data
  • Events All event inputs can be captured with waveforms
  • Trigger Point Pre- and post-trigger amount is user adjustable
  • Auto Arm Permits automatic stacking of captures
  • Auto Archive Automatic archive of capture to ZIP
  • Chart Playback all or any section at x1/8 to x8, auto
    playback supported
  • Ext. Sample Rates External TTL sample clock to 10,000 Hz
  • Review Formats Strip chart, numeric tabular, XY-Plot
  • Display Cursors Use cursors to select sections and make
    measurements
  • ZIP Archive Up to 45 million samples
  • ASCII conversion Direct conversion to PC/ASCII/EXCEL formats
Signal Inputs
  • Waveform Inputs 16 isolated, universal inputs
  • Waveform Isolation 250Vrms
  • Connector Banana (Voltage); Screw Terminal (Differential,
    Thermocouple and Bridge)
  • Max. Rated Input ±250Vrms (Voltage); ±40V (Differential,
    Thermocouple and Bridge)
  • Measuring Ranges 0.2Vfs to 400 Vfs (Voltage);
    5mVfs to 1600Vfs (Differential, Bridge);
    -100 to 1300°C (Thermocouple, depending on type)
  • Bandwidth 20 kHz (-3dB) (Voltage); 10 Hz (-3dB) (TC)
  • Excitation Isolated, adjustable to 10.5V@2.5mA
  • Input Impedance 1 Megohm
  • Filter Choices Low Pass with stops from 10 Hz to 20,000 Hz;
    edge; High pass with starts from 0.1 Hz to 100 Hz;
    Band Pass; Notch with 50 or 60 Hz center
  • RMS time constant Selectable from 0.02 to 2 seconds
  • Event Inputs 16 TTL or switch closure
  • Waveform Testing All active waveform channels monitored
    simultaneously
  • Trigger Window, slope/level
  • Trigger Function Alarm output, trigger capture, abort capture
ZIP Drive
  • Function Setup files, software upgrades and data transfer/
    archive
  • Menu Functions Format, Rename, Delete, Copy, Print (ASCII)
Miscellaneous
  • Power AC: 85 to 250 VAC at 50 or 60 Hz (auto select);
    DC: 11 to 18 VDC

  • Power Consumption 100 W typical, 200 W maximum
  • Battery Provides orderly shutdown without data loss in the
    event of a power loss
  • Case Description Aluminum, 17”L x 11.2”W x 6.6”H, 30 lbs.
  • Environmental 5°C to 40°C (40°F to 105°F), 10% to 95% RH,
    non-condensing
  • Utility Port Connections for start/stop, remote drive and trigger
    Indicators Trigger, armed, battery/power
  • Controls Full alpha-numeric keypad, soft keys, encoder wheel
  • Post processing Frequency spectrum (FFT)
  • Built in Help/Reports General, System Status, Chart Information
  • Multi-language support French and German

Please contact us for more information on the Astro-Med Dash 16U

Agilent
E4437B

Agilent E4437B
Yokogawa
WT1600B


Yokogawa WT1600B
Anritsu
MS2026A

Anritsu MS2026a
Megger
DLRO-200-115

Megger DLRO-200-115

Teseq
NSG437
Varian 956 Porta Test
Solar
7054-1

Chauvin Arnoux 6121
Chroma
63209
Chroma 63209
Privacy Notice Site Map Terms of Use
Copyright © 1996-2007  |  Advanced Test Equipment Rentals  |  All Rights Reserved