Fluke 199B
Dual Input Oscilloscope

Isolated Inputs A and B (Vertical)

  • Bandwidth, DC Coupled: 200 MHz (-3 dB)
  • Lower Frequency Limit, AC Coupled:
    • with 10:1 probe......<2 Hz (-3 dB)
    • direct (1:1) ..... <5 Hz (-3 dB)
  • Rise Time: 1.7 ns
  • Analog Bandwidth Limiters ..... 20 MHz and 10 kHz
  • Input Coupling.AC, DC
  • Polarity ..... Normal, Inverted
  • Sensitivity Ranges C versions:
    • with 10:1 probe..... 20 mV to 1000 V/div
    • direct (1:1) . 2 mV to 100 V/div
  • Sensitivity Ranges B versions:
    • with 10:1 probe..... 50 mV to 1000 V/div
    • direct (1:1) . 5 mV to 100 V/div
  • Trace Positioning Range... 4 divisions
  • Input Impedance on BNC:
    • DC Coupled..... 1 MW(±1 %)//15 pF (±2 pF)
  • Max. Input Voltage
    • with 10:1 probe 600 V CAT III; 1000 V CAT II
    • direct (1:1) . 300 V CAT III
  • Vertical Accuracy (1.5 % + 0.04 range/div)
  • 2 mV/div: ... (2.5 % + 0.08 range/div)
  • Digitizer Resolution: 8 bits, separate digitizer for each input

Horizontal

  • Maximum Time Base Speed: 5 ns/div
  • Minimum Time Base Speed (Scope Record): 2 min/div
  • Real Time Sampling Rate (for both inputs simultaneously)
    • 5 ns to 2 s /div ..... up to 2.5 GS/s
    • 5 s to 120 s/div ...... 20 MS/s
  • Record Length
    • Scope Record Mode:  >27000 points on each input
    • Scope Normal Mode: 1200 points on each input
    • Scope Glitch Capture Mode ...... 300 min/max pairs on each input
  • Glitch Detection: 5 s to 120 s/div
  • Waveform Display.. A, B, A+B, A-B, A*B, A vs B Normal, Average (2,4,8,64 x), Persistence
  • Time Base Accuracy ±(100 ppm + 0.04 div)

Trigger and Delay

  • Trigger Modes: Automatic, Edge, External, Video, Pulse Width
  • Trigger Delay..... up to +1200 divisions
  • Pre Trigger View one full screen length
  • Max. Delay ... 12 seconds

Automatic Connect-and-View Trigger

  • Source... A, B, EXT
  • Slope..Positive, Negative

Edge Trigger

  • Screen Update: Free Run, On Trigger, Single Shot
  • Source A, B, EXT
  • Slope . Positive, Negative
  • Trigger Level Control Range .... 4 divisions
  • Trigger Sensitivity A and B:
    • DC to 5 MHz at >5 mV/div .... 0.5 divisions
    • DC to 5 MHz at 2 mV/div and 5 mV/div ... 1 division
    • 200 MHz (FLUKE 199B) 1 division
    • 250 MHz (FLUKE 199B) ...... 2 division

Isolated External Trigger

  • Bandwidth: 10 kHz
  • Modes: Automatic, Edge
  • Trigger Levels (DC to 10 kHz): 120 mV, 1.2 V

Video Trigger

  • Standards ... PAL, PAL+, NTSC, SECAM
  • Modes.... Lines, Line Select, Field 1 or Field 2
  • Source ......A
  • Polarity ... Positive, Negative
  • Sensitivity ...... 0.7 division sync level
 

Pulse Width Trigger

  • Screen Update On Trigger, Single Shot
  • Trigger Conditions.....<T, >T, =T (±10 %), .T(±10 %)
  • Source.... A
  • Polarity .Positive or negative pulse
  • Pulse Time Adjustment Range .. 0.01 div. to 655 div. 
    • with a minimum of 300 ns (<T, >T) or 500 ns (=T, .T),
    • a maximum of 10 s,
    • and a resolution of 0.01 div. with a minimum of 50 ns

Continuous Auto Set

  • Modes:
    • Normal 15 Hz to max. bandwidth
    • Low Frequency.. 1 Hz to max. bandwidth
  • Minimum Amplitude A and B
    • DC to 1 MHz...... 10 mV
    • 1 MHz to max. bandwidth 20 mV

Automatic Capturing Scope Screens

  • Capacity 100 dual input scope Screens

Automatic Scope Measurements

General

  • Inputs .... A and B
  • DC Common Mode Rejection (CMRR).... >100 dB
  • AC Common Mode Rejection at 50, 60, or 400 Hz >60 dB

DC Voltage (VDC)

  • Maximum Voltage
    • with 10:1 probe... 1000 V
    • direct (1:1) .... 300 V
  • Maximum Resolution
    • with 10:1 probe...... 1 mV
    • direct (1:1) .. 100 µV
  • Full Scale Reading ...... 1100 counts
  • Accuracy at 5 s to 5 s/div ..... (1.5 % +5 counts)
    • 2 mV/div..... (1.5 % + 10 counts)
  • Normal Mode AC Rejection at 50 or 60 Hz ..... >60 dB

AC Voltage (VAC)

  • Maximum Voltage
    • with 10:1 probe.1000 V
    • direct (1:1) ..300 V
  • Maximum Resolution
    • with 10:1 probe....1 mV
    • direct (1:1) 100 µV
  • Full Scale Reading... 1100 counts
  • Accuracy DC coupled: DC to 60 Hz: (1.5 % +10 counts)
  • AC coupled, low frequencies:
    • 50 Hz direct (1:1) ..(2.1 % + 10 counts)
    • 60 Hz direct (1:1) ..(1.9 % + 10 counts)
  • AC or DC coupled, high frequencies:
    • 60 Hz to 20 kHz....(2.5 % + 15 counts)
    • 20 kHz to 1 MHz ......(5 % + 20 counts)
    • 1 MHz to 25 MHz...(10 % + 20 counts)
    • Normal Mode DC Rejection>50 dB

AC+DC Voltage (True RMS)

  • Maximum Voltage:
    • with 10:1 probe ... 1000 V
    • direct (1:1) ...... 300 V
  • Maximum Resolution
    • with 10:1 probe ...... 1 mV
    • direct (1:1) .... 100 µV
  • Full Scale Reading ... 1100 counts
  • Accuracy:
    • DC to 60 Hz .... (1.5 % + 10 counts)
    • 60 Hz to 20 kHz ...... (2.5 % + 15 counts)
    • 20 kHz to 1 MHz (5 % + 20 counts)
    • 1 MHz to 25 MHz ..... (10 % + 20 counts)

 

Please contact us for more information on the Fluke 199B


Haefely
ECOMPACT4

Haefely Ecompact 4
CA Instruments
5001ix


California Instruments 5001ix
Voltech
PM6000

Voltech PM6000
IFI
T4026-40

IFI T4026-40

Teseq
NSG437
Varian 956 Porta Test
Solar
2654-1

Solar 2654-1
Privacy Notice Site Map Terms of Use
 
Copyright © 1996-2008  |  Advanced Test Equipment Rentals  |  All Rights Reserved