DC Voltage Measurement
- Range: 30V, 20 V, 90 mV
- Resolution: 0.001V, 0.001 V, 0.01 mV
- Accuracy, (% of Reading + Counts): 0.02 % + 2
Voltage Source
- Range: 100 mV, 10 V
- Resolution: 0.01 mV, 0.001 V
- Accuracy, (% of Reading + Counts): 0.02 % + 2
- Maximum load: 1 mA
Millivolt Measurement and Source
- Range: -10 mV to 75 mV
- Resolution: 0.01 mV
- Accuracy: ±(0.025 % + 1 count)
- Maximum input voltage: 30 V
DC mA Measurement and Source
- Range: 24 mA
- Resolution: 0.001 mA
- Accuracy, (% of Reading + Counts): 0.02 % + 2
- Drive capability: 1000W at 20 mA
|
Ohms Measurement
- Ohms Range: 0 to 400 W,
400 to 1.5 kW, 1.5 to 3.2 kW
- 4-Wire: 0.1, 0.5, 1±W
- 2- and 3-Wire: 0.15, 1.0, 1.5 ±W
- Excitation Current: 0.2 mA
- Maximum input voltage: 30 V
Ohms Source
- Ohms Range: 15 to 400 W
- Excitation Current from Measurement Device: 0.15
to 0.5 mA
- Accuracy: ±0.15W
- Ohms Range: 15 to 400 W
- Excitation Current from Measurement Device: 0.5
to 2 mA
- Accuracy: ±0.1W
- Ohms Range: 400 to 1.5 kW
- Excitation Current from Measurement Device: 0.05
to 0.8 mA
- Accuracy: ±0.5W
- Ohms Range: 1.5 to 3.2 kW
- Excitation Current from Measurement Device: 0.05
to 0.4 mA
- Accuracy: ±1W
- Resolution:
- 15 to 400 W
- 400 to 3.2 kW
Frequency Measurement
- Range: 2.0 to 1000.0 CPM
- Resolution: 0.1 CPM
- Accuracy: ± (0.05 % + 1 count)
- Range: 1 to 1100
- Resolution: 0.1 Hz
- Accuracy: ± (0.05 % + 1 count)
- Range: 1.0 to 10.0 kHz
- Resolution: 0.01 kHz
- Accuracy: ± (0.05 % + 1 count)
- Sensitivity: 1V peak-to-peak minimum
- Waveform: squarewave
|