| Frequency Range |
% of Reading + Number
of Counts |
| 1 Year |
2 Year |
| 20 Hz to 40 Hz |
2% + 10 |
2% + 10 |
| 40 Hz to 500 Hz |
0.5% + 5 |
0.5% + 5 |
| 500 Hz to 1 kHz |
2% + 10 |
2% + 10 |
| 1 kHz to 5 kHz |
10% + 20 |
10% + 20 |
| Ranges :
1.1000 V, 11.000 V, 110.00 V, 300.0 V rms |
| Resolution:
11.000 counts in all ranges except 300 V; 3,000 counts on 300 V
range. |
| Input Impedance: 5
M W and <100 pF |
| Temperature Coefficient: 10%
of specification/°C in the ranges -10 to 18°C and 28 to 50°C |
| Input Coupling:
ac |
| Maximum Input Voltage:
300 V rms |
| Minimum Input Voltage:
0.5 V above 1 kHz |
| Specifications apply for 10% to 100%
of voltage range. |