Platform
CM-BASE:CEMaster base platform includes: built-in, 10A coupler/decoupler for AC and DC; external interlock, EUT output connection (NEMA 5-15, CEE7 "Schuko," or BS 1363 British Standard).
CM-SW:Easy to use Windows® application software with predefined IEC routines and the ability to run any sequence of tests in a single "Run" operation.
CM-FP:Front Panel keypad and graphics display for local operation and control with predefined IEC routines.
Optional Capabilities:
- ESD Testing
- EFT Testing
- Surge Testing
- Power Frequency Magnetic Field Testing
- Pulse Magnetic Field Testing
- Dips and Interrupts Testing
|
|
Keytek CEMaster EMC / EMI Immunity Test Simulator
The Keytek CEMASTER EMC / EMI Immunity Test Simulator is a customer configured tester that allows you to use a system specific to your needs and budget. It provides compliance-level testing to IEC 1000-4-X Series Immunity Standards. Starting with a base platform, you add only the functions outlined in your product test plan. If, in the future, your testing requirements change, your CEMASTER may be upgraded to incorporate the additional functions required to meet your testing parameters.
General Specifications
| Test Type |
Model |
Standard |
Feature |
Range |
| ESD |
CM-ESD |
IEC 1000-4-2 |
Peak Air Discharge |
500V-8.8kV |
| Peak Contact Discharge |
500V-4.4kV |
| Repetition Rates |
Single Shot, 1 pulse/sec and 20 pulses/sec |
| Waveform |
|
| EFT |
CM-EFT |
IEC 1000-4-4 |
Peak Voltage |
250V-2.5kV |
| Frequency |
1-100kHz |
| Waveform |
|
| Surge |
CM-SURGE |
IEC 1000-4-5 |
Voltage Waveform |
1.2/50µs |
| Peak Voltage |
250V-2.5kV |
| Current Waveform |
8/20µs |
| Peak Current |
125A-1.25kA |
| Waveform |
|
| Repetition Rate |
To 6 per minute at 500V |
| Power Frequency Magnetic Field |
CM-HPWR |
IEC 1000-4-8 |
Field Frequency |
50/60Hz |
| Magnetic Field |
0.5A/m - 4A/m |
| Pulse Magnetic Field |
CM- HPULSE |
IEC 1000-4-9 |
Field Pulse |
8/20µs |
| Amplitude Field |
50A/m - 800A/m |
| Test Parameters |
|
| Dips & Interrupts |
CM-PQF |
IEC 1000-4-11 |
Dips |
70%, 40% |
| Interrupts |
0% (Short or Open) |
| Transition Time |
1µs - 5µs |
| Inrush |
>250 Amps @ 120V |
| >500 Amps @ 240V |
| AC Mains Monitor |
0 - 265 VAC |
| Test Parameters |
|
| Power Requirements |
Input Voltage: |
90 - 250VAC |
| Input Current: |
1A @ 120V; 0.5 @ 240V |
| Coupler/Decoupler Range |
AC Voltage Ratings: |
50 - 250VAC |
| AC Current Ratings: |
10A continuous, >16A start-up |
| DC Voltage Ratings: |
0 - 100VDC |
| DC Current Ratings: |
10 Amps |
| Physical |
Dimensions: |
7.0" H x 18.5" W x 22" D |
| Approx. Weight: |
64 lbs |
|