The Rod-L Hipot Testers are designed to perform the tests specified by U.L, I.E.C., C.S.A., V.D.E. and F.C.C.
Voltage and Current Configurations
| Maximum Output Voltage |
Maximum Current Limit |
| 10mA |
25mA |
40mA |
50mA |
70mA |
| 1000 VAC |
x |
x |
x |
x |
x |
| 1500 VAC |
x |
x |
x |
x |
|
| 2800 VAC |
x |
x |
x |
|
|
| 5000 VAC |
x |
x |
|
|
|
Standard Features
| |
Models |
| AVS5 |
BVS5 |
| Variable Voltage |
Yes |
Yes |
| Variable Current Limit |
Yes |
Yes |
| Real Current Detector |
No |
Yes |
| Front Panel Dwell Time Select |
Yes |
Yes |
| Front Panel Rise Time Select |
Yes |
Yes |
Specifications
| Input Power |
115/230 VAC, 50/60 Hz, 300-watts for M100 |
| Test Voltage |
User specified maximum - adjustable from zero volts to full scale |
| Voltage Ramp Time |
1 second to 30 seconds, user adjustable |
| Test Interval |
1 to 90 seconds, user adjustable |
| Shutdown |
2 milliseconds after failure |
| Weight |
26 lbs Net; 30 lbs Shipping |
|