Solar 6254-5S RFI Transient Generator
for conducted transient susceptibility testing up to 250 volts, peak
The Model 6254-5S RFI Transient Generator
provides up to 250 volts peak amplitude. The output transient shape
follows the curve given in Figure 19 of MIL-STD-462. Less than 1.0
microsecond rise time, falling to zero in 8 to 14 microseconds, crossing
through zero to "ring" in the manner of an inductive transient and
returning to zero again as it "rings." The amplitude of the transient
is adjustable from less than 10 volts to over 250 volts peak. Using series
injection on 50, 60 or 400 Hz lines, the transient can be applied to the
positive or the negative half-cycle. The transient's relation to the
sine wave may be adjusted in phase from 0° to 360°. For non-synchronous
injection, the repetitive rate of the transient can be adjusted from 0.5
to 500 p.p.s.
The Model 6254-5S RFI Transient Generator
was especially designed for screen room use in making conducted transient
susceptibility tests as required by military specifications. These
specifications include: parts of MIL-STD-461A/462, MSFC-STD-279, Lockheed
422966 (L1011), TRW TOR-1001,Douglas WZZ-7000 (DC-10), and others.
Features:
-
Provides outputs from less than
10 volts to over 250 volts peak amplitude into high impedance loads
and more than 35 kw into 0.5 ohm load.
-
Wider range of repetition rates
allows greater utilization in empirical setups.
-
Output terminals for series or
parallel injection.
-
Single pulse feature enables
controlled isolation of transient effects.
-
Adjustable pulse position on a.c.
lines relates the transient susceptibility to the real time aspects of
digital systems.
-
Transient may be injected in
synchronism with repetitive circuit functions as required by Method CS06
of MIL-STD-462.
-
Remote triggering of individual or
repetitive pulses in terms of particular system characteristics.
Specifications:
-
Spike Amplitude: Continuously
adjustable from less than 10 volts to more than 250 volts peak.
-
Repetition Rate: Continuously
adjustable from 0.5 to 500 p.p.s.
-
Rise Time: Less
than 1.0 microsecond.
-
Spike Duration: Output
falls to zero in approximately 8 to 14 microseconds.
-
Spike Shape: Ringing
characteristic as shown in Figure 19 of MIL-STD-462.
-
Phase Adjustment: Spike
position adjustable from 0° to 360° on 50, 60 or 400 Hz lines.
-
Sync Operation: Triggers
at any rate from once every ten seconds to over 800 transients per
second.
-
Internal Impedance: 0.5
ohm.
-
Output Power: More
than 35 kw peak into 0.5 ohm load.
-
Power Current in Series Injection Mode: Handles up to 50 amperes at power frequencies.
-
Power Requirements: 115
volts 60 Hz, 1.8 amperes. (230 volts 50 Hz, 0.9 ampere available.)
-
Size: 8.125"
wide, 9" high, 14.625" deep. (20.64 cm x 22.86 cm x 37.15 cm.)
|