Tektronix DPO4104 1 GHz Digital Oscilloscope
Offering up to 1 GHz bandwidth and 5 GS/s sample rate, the Tektronix DPO4104 Mixed Signal Oscilloscope Series delivers the performance you need to see fast-changing signal details. With as many as 20 channels, you can analyze analog and digital signals with a single instrument. Combine that with automated serial and parallel bus analysis, innovative Wave Inspector® controls, and automated power measurements, and the DPO4104 provides the feature-rich tools you need to simplify and speed debug of your complex design.
Tektronix DPO4104 Features:
- 1 GHz, 500, 350 MHz Bandwidth Models
- 2 and 4 Channel Models
- Sample Rates Up to 5 GS/s on All Channels
- 10 Megasample Record Length on All Channels
- Wave Inspector Controls Provide
Unprecedented
Efficiency in
Waveform Analysis
- I2C, SPI, CAN Serial Triggering
and Analysis
- 10.4 in. (264 mm) XGA Color Display
- Small Footprint and Lightweight –
Only 5.4 in.
(137 mm) Deep and
11 lbs. (5 kg)
- USB and CompactFlash on Front
Panel for Quick
and Easy Storage
- Built-in Ethernet Port
- Plug ‘n’ Play Connectivity and
Analysis
Software Solutions
- USB 2.0 Device Port for Direct
PC Control of
Oscilloscope
Using USBTMC
- Suite of Advanced Triggers
- e*Scope® Remote Viewing
and Control
- Interoperability with Tektronix
Logic Analyzers
- TekVPI™ Probe Interface Supports
Active,
Differential and Current
Probes for Automatic
Scaling
and Units
Tektronix DPO4104 Applications:
- Embedded Design and Debug
- Investigation of Transient
Phenomena
- Power Measurements
- Video Design and Debug
- Spectral Analysis
- Automotive Electronics Design
and Debug
- Manufacturing Test and
Quality Control
- Electro-mechanical Design
and Analysis
- Bio-medical Product Development
- Industrial Control
Vertical System
| Input Channels |
4 |
| Analog Bandwidth (–3 dB) |
1 GHz |
|
Calculated Rise Time |
350 ps |
|
Hardware Bandwidth Limits |
20 MHz or 250 MHz |
| Input Coupling |
AC, DC, GND |
| Input Impedance |
1 MΩ ±1%, 50 Ω ±1% |
|
Input Sensitivity, 1 MΩ |
1 mV/div to 10 V/div |
| Input Sensitivity, 50 Ω |
1 mV/div to 1 V/div |
|
Vertical Resolution |
8 bits |
| Max Input Voltage, 1 MΩ |
250 VRMS with peaks ≤±400 V |
| Max Input Voltage, 50 Ω |
5 VRMS with peaks ≤±20 V |
| DC Gain Accuracy |
±1.5% with offset set to 0 V |
| Offset Range |
1 mV/div to 50 mV/div ±1 V
50.5 mV/div to 99.5 mV/div ±0.5 V
100 mV/div to 500 mV/div ±10 V
505 mV/div to 995 mV/div ±5 V
1 V/div to 5 V/div ±100 V
5.05 V/div to 10 V/div ±50 V |
|
Channel-to-Channel Isolation |
≥100:1 at ≤100 MHz and ≥30:1 at >100 MHz up to the rated bandwidth
(Any Two Channels at Equal
Vertical Scale) |
Horizontal System
| Maximum Sample Rate (all channels) |
5 GS/s |
| Maximum Record Length
(all channels) |
10 M points |
| Maximum Duration at Highest Sample Rate (all channels) |
2 ms |
| Timebase Range (s/div) |
400 ps to 1,000 s |
| Timebase Delay Time Range |
–10 divisions to 50 s |
| Channel-to-channel Deskew Range |
±100 ns |
| Long Term Sample Rate and Delay Time Accuracy |
±5 ppm over any ≥1 ms interval |
| Delta Time Measurement Accuracy |
±(1/sample rate + 5 ppm x |Reading| + 0.4 ns) |
|