Power Requirements

  • Standard (ac power): 90 to 132 Vac or 198 to 264 Vac, automatically selected; 50 to 60 Hz
  • Option AWQ: +19 Vdc to +59 Vdc ( >+23 Vdc required to start )

GPS Antennas and Antenna Assemblies

  • HP 58504A GPS L1 antenna
  • HP 58510A environmental cover and ground plane
  • HP 58513A antenna assembly - integrates a stainless steel mast and ground plane with the antenna, antenna cable and environmental cover
  • HP 58538A/ 58539A lightning arrestors; provides protection against nearby lightning strikes.

 

GPS Time and Frequency Reference Receiver

The HP / Agilent / Symmetricom 58503B GPS Time & Frequency Reference Receiver is a high quality source of precision time, time interval, and frequency in a lightweight compact package. The high-performance 58503B ensures excellent accuracy and reliability— the 58503B delivers precise time and frequency virtually anywhere in the world at an affordable price. When locked to the GPS signal, the Agilent / HP 58503B provides a 10 MHz signal with an accuracy of better than 1x10–12 using a 1 day average. A 1 pulse per second (1 PPS) signal with timing jitter of <750 ps rms is provided. Time accuracy relative to GPS of 20ns (1 Sigma, typical) can be obtained once the unit is properly installed and all systematic offsets have been compensated out. If the GPS signal is interrupted, the Agilent / HP 58503B enters an intelligent holdover mode, which ensures frequency is maintained to better than 1x10–10 per day (phase accumulation <8.6 µsec after 24 hours).

Features:

  • Reduce the cost for precise time and frequency
  • Eliminate calibration through GPS technology
  • Improve reliability through HP quartz oscillators

GPS Receiver

  • 8-channel, parallel-tracking
  • HP  SmartClock Enhanced GPS
  • DC power options available

10 MHz Output Characterizations

  • Frequency Accuracy (Locked): < 1 x 10¯²² for a one day average
  • Holdover Aging (Unlocked): < 1 x 10¯²° per-day average frequency change in 24 hours of unlocked operation
  • Output Level: > 1V p-p sine wave into 50 ohm load

1 pps Output Characteristics

  • Jitter on Leading Edge (Locked): < 750ps rms
  • Time Accuracy (Locked): <110ns with respect to UTC (USNO MC) - 95% probability when unit is properly installed and calibrated
  • Accumulated Time Error (Unlocked): < 8.6 µs / accumulated in 24 hours after three days of locked operation with a fixed antenna location
  • Output Level: > 2.4V pulse into 50 ohm load
  • Pulse Width: 26 µs

Please contact us for more information on the Agilent / HP 58503B GPS Time and Frequency Reference Receiver


Featured Test Equipment:

Rent an Haefely ECompact 4 Transient Immunity Test Set

Haefely ECOMPACT4
Transient Immunity Test

Voltech PM6000
Voltech PM6000
Power Analyzer



IFI T4026-40
TWT Amplifier

Varian 956 Porta Test
Teseq NSG 438
ESD Gun

Solar 2654-1
Solar 2654
Transient Generator