| Power
Input 120 Vac at 50 or 60Hz Standard (must be specified) |
| Recommended
source: |
NEC
15 Ampere 120V single phase. |
| Recommended
ground: |
Solid metallic with a resistance of
less than 5 ohms. |
| Voltage
(rms): |
120V:
Nominal rating,
102 to 132V: Operating range. |
| Current
maximum (rms): |
9 Amperes continuous on BURN mode.
9 Amperes pulsing in IMPULSE mode. |
| Optional
Power Input, 220 or 240V at 50 or 60 Hz (must be specified.) |
| Recommended
source: |
NEC
15 Ampere 240V single phase. |
| Recommended
ground: |
Solid
metallic with a resistance of less than 5 ohms. |
| Voltage
(rms): |
240V
full rating,
204 to 264V operating range or 220V full rating,
198V to 242V operating range. |
| Current
maximum (rms): |
5 Amperes continuous in BURN mode.
5 Amperes pulsing in IMPULSE mode. |
| Output |
| |
Impulse
Mode |
| |
Peak
Voltage: |
25
kV maximum, continuously variable 0 to 25 kV. |
| |
Capacitance: |
4
μF. |
| |
Stored
Energy: |
1250
Joules (watt-secs) at 25 kV (maximum). |
| |
Peak
Current: |
Approximately
16 kA (50 ft. output cable short-circuited). |
| |
Time
for Voltage
Pulse to decay to 5%: |
150
microseconds (50 ft . output short-circuited). |
| |
Polarity: |
Negative
with respect to ground. |
| |
Impulse
Timing: |
Fixed.
6 seconds for 60 Hz;
7.2 seconds for 50 Hz. |
| |
Internal
Energy
Discharge Time: |
Less
than 10 seconds to 1% of starting value (4 pF). |
| |
Test
Energy
Discharge Time: |
Less
than 10 seconds to 1% of starting value for a test cable capacitance
of 4 uF (time porportional to load capacitance). |
| |
Proof/Burn
Mode: |
| |
Peak
Voltage: |
30
kV maximum, continuously variable 0 to 30 kV. |
| |
Polarity: |
Negative
with respect to ground. |
| |
Voltage
Build-Up Time: |
5
kV per second initial rate for 4 pF test cable. |
| |
Discharge
Time: |
Less
than 10 seconds to 1% of starting value for a test cable capacitance
of 4 pF. Time proportional to load capacitance. |
| |
Peak
Current: |
50
mA continuous 50 or 60 Hz into short circuit. |