| Control Method |
| Method |
Continuous control of set temperature and humidity. Program operation. Temperature heat-up or pull-down rate. Humidity control (M type). |
| Temperature & humidity setting control action |
Direct setting of temperature and relative humidity PID control. SSR drive system. |
| Power Supply |
| 50/60 Hz Single Phase |
| Total Loaded Current |
| 200V AC |
20 Amps |
| 220V AC |
18.5 Amps |
| Performance |
| Unsaturated control |
| Temperature control range |
+105.0 to +162.2°C |
| Relative humidity control range |
75 to 100% RH |
| Pressure control range |
0.2 to 4.0 kg/cm²G |
| Temperature & humidity constancy |
±0.5°C/±3% RH |
| Temperature uniformity |
±0.7°C (at 75% RH) |
| Heat up and pressurizing time (at RT +23°C) |
0 to 4.0 kg/cm²G About 90 minutes |
| Saturated control |
| Temperature control range |
+105.0 to +151.1°C |
| Pressure control range |
0.2 to 4.0 kg/cm²G |
| Temperature constancy |
+0.5°C |
| Temperature uniformity |
+0.5°C |
| Heat up and pressurizing time (at RT +23°C) |
0 to 4.0 kg/cm²G About 90 minutes |
| Dry and wet-bulb temperature control |
| Temperature rise |
| Temperature control range |
+105.6 to +162.2°C |
| Relative humidity control range |
75 to 95% RH |
| Heat up and pressurizing time (at RT +23°C) |
0.2 to 4.0 kg/cm²G About 120 minutes |
| Test process |
| Temperature control range: |
+105.6 to +162.2°C |
| Relative humidity control range |
75 to 98% RH |
| Pressure control range |
0.2 to 4.0 kg/cm²G |
| Temperature & humidity constancy |
±0.5°C/±3% RH |
| Temperature uniformity |
±0.7°C (at 75% RH) |
| Temperature pull-down |
| Temperature control range |
+50.0 to +95.0°C |
| Temperature pull-down rate (at RT +23°C no-specimen) |
From: +162.2°C/75% RH To: +85.0°C/85% RH) Min. about 120 minutes |
| Hold process |
| Temperature control range |
+50.0 to +95.0°C |
| Relative humidity control range |
75 to 95% RH |
| Cloth Wick |
200 hours continuous use (with no specimen) |