Configuration shown includes Partlow 1462 ramping controller with RS232 interface and window (also available with Watlow 942 Controller)
|
|
Tenney Compact Junior Line of Temperature Chambers
Tenney 's compact junior line of chambers in both
bench-top and floor models are capable of simulating a wide range of
temperature conditions. These chambers are well-suited for use in
electronic, military, and pharmaceutical quality assurance and reliability
testing, as well as research testing and production processes. In keeping
with the needs of today 's lab, we specifically designed these chambers
to have a compact exterior, yet an ample interior workspace to maximize
valuable floor space. Available in with ramping
controller, RS232 interface, RS485 interface, window.
Specifications
| Model |
TJR |
| Workspace |
W |
16" |
| D |
11" |
| H |
12" |
| Exterior |
W |
37" |
| D |
22-1/2" |
| H |
30-5/8" |
| Temperature
Range °C |
| Low |
-75 |
| High |
+200 |
Change
Rates in Minutes
Ambient to °C to Chamber Empty |
| 200° |
60 |
| 185° |
50 |
| 160° |
40 |
| 140° |
30 |
| 105° |
20 |
| 70° |
10 |
| 25° |
0 |
| -15° |
10 |
| -40° |
20 |
| -54° |
30 |
| -65° |
40 |
| -73° |
55 |
| -75° |
Ultiimate |
Live
Load Capacity in Watts (humidity system off)
Temperature °C |
| -40° |
170 |
-54° |
145 |
-65° |
105 |
-73° |
60 |
-75° |
Ultimate |
| Utilities,
etc |
| Refrigeration |
(2)
1/2 HP |
| Heater
Capacity |
500
W |
| Amps@120V
/ 1~ / 60Hz |
18 |
| Amps
Fuse |
20 |
| Unit
Weight - lbs/kg |
292
/ 132 |
|