Associated Research 3140 40 Amp Ground Bond Tester

Associated Research 3140 40 Amp Ground Bond Tester
Request Quote Print Page
Datasheet

Features:

  • Patented VERI-CHEK feature prompts users through steps to validate the instrument's operation
  • Graphic LCD provides intuitive menu system to simplify the entire testing process from set-up to results
  • Patented CAL-ALERT feature alerts the operator that the tester is due for re-calibration
  • PLC Remote Control allows operators to remotely control the Ground Bond tester
  • 10 Memories with 3 steps per memory for storing and recalling test parameters
  • Interconnects with an Associated Research Hipot tester to form a complete test system
  • 4 wire measurement and milli ohms offset for accurate Ground Bond test results
  • Electronic Dwell timer for more consistent and reliable testing

 

Specifications:

Input
Voltage 115/230 VAC ± 10%, user selectable
Frequency 50/60 Hz ± 5%
Fuse - 3130 6.3 A, slo-blow 250 VAC
Fuse - 3140 10 A, slo-blow 250 VAC
Fuse - 3130 15 A, slo-blow 250 VAC
Ground Bond Test Mode
Output Rating Current: 1.00 - 40.00 A AC
Voltage: 8 VAC, fixed
Resolution: 0.01 A
Regulation: ± (2% of setting + 0.02 A)
Output Frequency Range: 50 / 60 HZ, User Selectable
Dwell Time Setting Range: 0 and 0.5 - 999.9 secs
0 for continuous running
Resolution: 0.1 sec
Accuracy: ± (0.1% of setting + 0.05 secs)
Max. & Min. Limits Range: 0 - 150 m ohms for 30.01 - 40.00 A
0 - 200 m ohms for 10.01 - 30.00 A
0 - 600 m ohms for 1.00 - 10.00 A
Accuracy: ± (3% of setting + 3 m ohms)
Offset Capability Range: 0 - 100 mΩ
Resolution: 1 mΩ
Accuracy: ± (2% of setting + 2 mΩ)