Features
-
Broad Wavelength Response 1100 to 1700 nm
-
High Bandwidth DC up to 1.2 GHz
-
High Gain 1 V/mW
-
Probe Connects Directly to TDS Series Scope (TEKPROBE®) or Other 50 Ohm Instruments with 1103 TEKPROBE Power Supply
-
SONET/SDH up to 622 Mb/s
Applications
-
Eye-pattern Testing of Optical Communication Signals (SONET/SDH and Fiber Channel)
Characteristics
The optical input is a one meter, 62.5 µm multimode fiber with an FC/PC connector.
-
Wavelength response: 1100 to 1700 nm
-
Bandwidth*1 (Typical): DC to 1.2 GHz
-
Conversion Gain: 1 V/mW
-
Max. Input Optical Power: 1 mW (0 dBm)*2
-
Max. Input Fiber Core Diameter: 62.5 µm
*1Optical bandwidth (-6 dB electrical)
*2Maximum average operating power