Vitrek 951i 6KV AC/DC HiPot, IR & Low Resistance Tester

Vitrek 951i 6KV AC/DC HiPot, Insulation Resistance & Low Resistance Tester
Request Quote Print Page
Datasheet

Features:

  • High Power Output
  • Wide Range of built-in Voltage Capabilities
  • Expansive Test Sequence Memory
  • Ground Bond Test Capability
  • 4-Wire Milli-Ohmmeter Function
  • Built-in Phase Angle Measurement
  • Multi-Dwell Functionality
  • Ramp High/Dwell Low Current Limits
  • Pico-Amp Leakage Measurement
  • Multi-Mode IR
  • Test Specific Fixture & Cable Compensation
  • Continuously Variable IR Test Voltage
  • Capacitance Test Modes
  • Light Weight Switching Power Supply Design

 

Specifications:

Line Leakage Current & Voltage Measurement
Voltage Measurement: 0 to ± 8KVDC 6KV RMS AC
Resolution: 0.1V, 1 V above 999.9V
DC Accuracy: 0.25% + 0.5V
AC Accuracy: 0.5% + 1.5V
Leakage Current: 0 to ± 200mA DC or RMS AC
Resolution: 4 digits (9999 counts) down to100pico-amps
DC Accuracy: 0.25% + 0.5nA
AC Accuracy: 0.5% + 20nA
Test Results Test Time: 0.02 to 9999 sec Last, Minimum, Average & Max V& A reading plus arc current
Pulse Mode (Flash) Test
Test Waveform: Trapezoidal (Selectable positive polarity, negative polarity or bi-polar)
Ramp Up/Down Time: 1ms (0.5ms for option AC2) to 30mS with 0.1ms resolution
Dwell Time: Resolution 1ms (0.5ms for option AC2) to 30mS with 0.1ms
Test Voltage: 50V to 8000V (20V to 2750V with option AC-2)
Resolution: 0.1V up to 999.9V, 1V above
Accuracy: 0.25% + 1.5V
Operating Temperature: 0 C to 50° C
Humidity: 90% RH, 0 to 40° C
Power: 115/230 VAC + 10%, 350 VA Max
Dimensions: 89mm H x 432mm W x 457mm D (3.5 " H x 17 " W x 18 " D)
Weight: 9Kg (18 lb.)