Primary Pump Type |
Dry Scroll - TS-620 |
Pump Rate |
30 m3/hr (500 l/m) |
Clean Room Use |
Yes |
Minimum Detectable Leak at 1000 ppm ambient He |
5 x 10-12 atm - cc / sec |
Maximum Test Port Pressure |
13 mbar, 10 Torr, 1330 Pa |
Helium Pumping Speed @ Test Port (Fine Test) |
1.8 l/s |
Calibration Routine |
Automated or Manual (Internal or External) |
Background Suppression |
Push Button Initiated Auto Zero and Auto Zero < Zero function |
User Interface |
High Clarity, Color Display, TFT Touch Screen |
Selectable Languages |
English, French, German, Japanese, Korean, Mandarin, Spanish |
Automated Cycling |
Programmable Rough Time, Test Time, Reject Set Points |
Response Time |
< 0.5 seconds |
Set Points |
Standard, 5 Set points, N/O or N/C; 3 Leak Rate, 1 pressure, 1 Audio |
Serial Interface |
RS-232 optically-isolated interface @ 9600 baud (DB-25S) |
Discrete I/O |
Optically isolated, 5-24 volts VDC outputs and inputs (DB-25S) |
Input Voltage |
120 Volts |