Safety:
- Door safety switch protects operator by removing power to the heating elements upon opening the door
- Thermocouple break protection cuts power to the heating elements, prevents a thermocouple failure runaway condition
- The rear of the chamber incorporates a 0.38" (0.95 cm) diameter port for monitoring chamber temperatures with independent measuring devices
Features
- Ceramic fiber insulation permits faster heat-up, reduces energy consumption.
- Embedded heating element on top and both sides improves temperature uniformity
- Drop-down door doubles as a convenient shelf for loading and unloading
- Operation: Digital single set point control. A single display shows actual temperature or set point
- 1100°C maximum temperature
- 76 cubic inch capacity
Specifications
Amps | 8.9 |
Chamber Cubic Inches L | 76 (1.3) |
Chamber Dimensions D | 5.0 (13) |
Chamber Dimensions H | 3.8 (9.8) |
Chamber Dimensions W | 4 (10.3) |
Hertz | 50/60 |
Operating Temp Range C | 100 - 1100 |
Overall Dimensions D | 13 (33) |
Overall Dimensions H | 14 (36) |
Overall Dimensions W | 9 (23) |
Stability at 1000 | ± 0.3°C |
Sub Product Group Id | 95 |
Temp Control Group | A1 |
Uniformity at 1000 | ±7.8°C |
Volts | 120 |
Watts | 1060 |
Applications
- Ashing most types of organic, inorganic samples
- Heat treating small steel parts
- Ignition tests
- Gravimetric analysis
- Determination of volatile and suspended solids