Boonton 4200 Power Meter, RF Microwattmete

Boonton 4200 Power Meter, RF Microwattmete
Enlarge Image
Boonton 4200 Power Meter, RF Microwattmete Boonton 4200 Power Meter, RF Microwattmete Boonton 4200 Power Meter, RF Microwattmete
Boonton 4200
Request Quote Print Page
Manual Datasheet
100 kHz - 100 GHz, -70 dBm to +37 dBm Dual Sensor Power Meter
 
Options: 

(please consult your sales representative on availability)

  • -01B IEEE-488 Bus Interface
  • -03A Second Input Channel

Boonton 4200 Power Meter, RF Microwattmeter

 
Boonton 4200 Features
  • Frequency Range, 100 kHz to 100 GHz (sensor dependent)
  • Power Range 0.1 nW (-70 dBm) to 5 W (+37 dBm) sensor dependent
  • Dual Sensor Simultaneous Power Measurements (optional)
  • 20 Measurements per Second
  • Stored Calibration Data for up to 8 Sensors
Boonton 4200 Specs
  • Frequency Range: 100 kHz to 100 GHz, sensor dependent. Refer to the Power Meter Sensor section
  • Power Range: (Display calibrated in mW, WW, nW, dBm, and dB relative to selected reference); -70 dBm to +37 dBm (0.1 nW to 5W), sensor dependent
  •  
Power Reference:
  • Source: Internal 50 MHz oscillator with Type N female connector on front panel
  • Power Output: 1.00 mW
  • Calibration Factors: 3.0 dB to -3.0 dB range in 0.01 dB steps
  • Ranging: Autoranging plus hold on range
  • Setting Time: Diode and millimeter wave sensors, typically 0.2 to 0.5s except 1.35 to 2.5s on the lowest range.  Thermocouple sensors, typically 0.5 to 6s
  •  
Recorder Output:
  • Watt Mode: 10 volts full-scale, proportional to indicated power over each range
  • dB Mode: 8 volts equivalent to 0 dBm. Sensitivity of 1 volt per 10 dB change over the entire dB range of sensor
  • Display: 4 digit LED, 3½ digit display of power, 4 digit display of dB with 0.01 dB resolution. Auxiliary analog display, uncalibrated, proportional to recorder output
  • Limits: Entered through front panel in dB only, operable in both dB and power modes
  • Annunciators: LEDs indicate mW, µW, nW, dBm or relative dB (dBr); use of channel 1 (CH1), channel 2 (CH2, option -03) and channel 3 (CH3 = CH1 - CH2 in dB); out of dB limits; and condition of GPIB activity (LSN, ATN, REM, TLK, option -01B)

Power Sensors

Diode Sensors:
Model
(Impedance)
(RF Connector)
Frequency Range Dynamic Range (dBm) Overload Rating
(Watts)
(dBm)
51011 (4B)
50 ohm
N(M)
100 kHz to 12.4 GHz -60 to +10 300 mW
+25 dBm
51012 (4C)
75 ohm
N(M)
100 kHz to 1 GHz -60 to +10 300 mW
+25 dBm
51012-S/4 (4C)
75 ohm
N(M)
100 kHz to 2 GHz -60 to +10 300 mW
+25 dBm
51013 (4E)
50 ohm
N(M)
100 kHz to 18 GHz -60 to +10 300 mW
+25 dBm
51075
50 ohm
N(M)
100 kHz to 18 GHz -60 to +10 300 mW
+25 dBm
51051 (4G)
50 ohm
APC 3.5 (M)
1 MHz to 26.5 GHz -60 to +10 300 mW
+25 dBm
51015 (5E)
50 ohm
N(M)
100 kHz to 18 GHz -50 to +20 2 W AVG.
+33 dBm
51033 (6E)
50 ohm
N(M)
100 kHz to 18 GHz -40 to +30 2 W AVG.
+33 dBm
51071
50 ohm
K(M)
10 MHz to 26.5 GHz -60 to +10 300 mW
+25 dBm
51072
50 ohm
K(M)
30 MHz to 40 GHz -60 to +10 300 mW
+25 dBm


Thermocouple Sensors:
Model
(Impedance)
(RF Connector)
Frequency Range Dynamic Range (dBm) Overload Rating
(Watts)
(dBm)
51100 (9E)
50 ohm
N(M)
10 MHz to 18 GHz -20 to +20 300 mW
+25 dBm
51101
50 ohm
N(M)
100 kHz to 4.2 GHz -20 to +20 300 mW
+25 dBm
51102
50 ohm
K(M)
50 MHz to 26.5 GHz -20 to +20 300 mW
+25 dBm