Fluke 1586A Super-DAQ Precision Temperature Scanner

Rent Fluke 1586A Super-DAQ Precision Temperature Scanner
Enlarge Image
Fluke Calibration 1586A
Request Quote Print Page
Datasheet
The Fluke 1586A Super-DAQ Precision Temperature Scanner is the most accurate and flexible temperature data acquisition system on the market. It scans and records temperature, dc voltage, dc current, and resistance for up to 40 input channels and scan speeds as fast as 10 channels per second. The Super-DAQ can be configured for use as a multi-channel data logger in the factory or as a precision reference thermometer for benchtop sensor calibration in the lab.
 
Fluke Calibration 1586A Features
  • Measure thermocouples, PRTs, thermistors, dc voltage, dc current, and resistance
  • Best-in-class temperature measurement accuracy:
    • PRTs: ± 0.005 °C (using external DAQ-STAQ Multiplexer)
    • Thermocouples: ± 0.5 °C (using High-Capacity Module and internal CJC)
    • Thermistors: ± 0.002 °C
  • Input Channels: Up to 40 isolated universal inputs
  • Flexible configuration: Internal High-Capacity Module and/or DAQ-STAQ Multiplexer
  • Selectable scan speed: Up to 10 channels per second
  • Four modes of operation: Scan, Monitor, Measure, Digital Multimeter (DMM)
  • Real-time color trending: Chart up to four channels simultaneously
  • Automated sensor calibration: Control Fluke Calibration temperature sources such as dry-wells or micro-baths for automated calibration routines
  • Data storage: Records up to 20 MB of data and setup files to internal non-volatile memory or to an external USB drive. Transfer data to a PC using the USB drive or LAN connection and view data in Microsoft® Excel
  • Data security: Administrator and user profiles for protecting settings and ensuring test traceability
  • Mx + B scaling and channel offset zero function
  • Alarms: Two independent, user-defined alarms for each channel indicate when an upper or lower range has been exceeded
Fluke Calibration 1586A Specs
Maximum Input: 50 V
Offset Voltage: <2 μV
3-Wire Internal Resistance Mismatch: <50 mΩ
Basic CJC Accuracy: 0.25°C