Fluke 39 Power Meter Harmonics Analyzer 1 Ch, 2.1 kHz

Fluke 39 Power Meter Harmonics Analyzer
Enlarge Image
Fluke 39
Request Quote Print Page
Manual Datasheet Other1 Other2

Specifications:

Voltage Range & Resolution: 5.0V to 600V rms ±5.0V to ±933V peak (AC+DC);
Accuracy: rms: ±(0.5% + 2 digits); Peak or DC: ±(2% + 3 digits) (Add 2 digits if < 15V rms)
Current (1mV/A) Isolated input Range & Resolution: 1.00A to 1000A rms (AC+DC); ±1.0A to ±2000A peak
Accuracy: rms: ±(0.5% + 3 digits)+probe specs; Peak or DC: ±(2% + 4 digits) + probe specs
Watts/Volt-Amps (1mV/A) isolated input Range & Resolution: 0.0W(VA) to 600kW(kVA) average; 0.0W(VA) to ±2000kW(kVA) peak
Accuracy: AC+DC: ±(1% + 4 digits) + probe specs
Harmonics (harmonic level >5% using Smooth 20) Volts: Fundamental to 13th ±(2%+2 digits); At 31st ±(8%+2 digits);
Amps or Watts: Fundamental to 13th ±(3% + 3 digits) + probe specs; At 31st ±(8% + 3 digits) + probe specs
Frequency Range & Resolution: 6.0 Hz to 99.9 Hz
Accuracy: ±0.3 Hz
Input Bandwidth DC, 6 Hz to 2.1 kHz
Crest Factor (CF) Range & Resolution: 1.00 to 5.00
Accuracy : ±4%
Power Factor (PF) Range & Resolution: 0.00 to 1.00
Accuracy ±0.02
Displacement Power Factor (DPF) Range & Resolution: 0.00 to 1.00
Accuracy: ±0.04 to ±0.03 (0.30 to 0.89) ±0.02 (0.90 to 1.00)
Phase Range & Resolution: -179º to 180º
Accuracy (Fundamental): ±2º + probe specs
K- Factor (KF) Range & Resolution: 1.0 to 30.00
Accuracy: ±10%
% THD-F Range & Resolution: 0.00% to 799.9%
Accuracy: ±(0.03 Reading + 2.0%)
% THD-R Range & Resolution: 0.0% to 99.9%
Accuracy: ±(0.03 Reading + 2.0%)